An abrupt rise of coagulation error messages on ACL TOP automated analysers

INTRODUCTIONBlood coagulation tests (BCT) are very important for clinicians to diagnose bleeding or thrombotic disorders and to monitor anticoagulant therapy. CASE DESCRIPTIONOn a Saturday morning, a laboratory technician noted an abrupt rise in the number of coagulation error messages on our ALC TO...

Full description

Saved in:
Bibliographic Details
Published in:Biochemia medica Vol. 29; no. 2; p. 021002
Main Authors: Calcoen, Bas, Desmet, Koen, Vermeersch, Pieter
Format: Report
Language:English
Published: 15-06-2019
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:INTRODUCTIONBlood coagulation tests (BCT) are very important for clinicians to diagnose bleeding or thrombotic disorders and to monitor anticoagulant therapy. CASE DESCRIPTIONOn a Saturday morning, a laboratory technician noted an abrupt rise in the number of coagulation error messages on our ALC TOP analysers. Visual inspection revealed the presence of partially and/or fully clotted citrate tubes and prompted the clinical biologist to further investigate a potential preanalytical cause. CONSIDERED CAUSESPartially or fully clotted blood in citrate tubes can have multiple causes including improper mixing of the tube, under- or overfilling or combining blood samples from different tubes into one citrate tube. WHAT HAPPENEDThe affected citrate tubes originated mostly from the same clinical departments. Moreover, all the affected tubes shared the same lot number (1 of 7 in use at the time). Visual inspection of 7 unopened boxes of 100 citrate tubes of this lot number revealed one box with nine completely empty and two partially filled tubes and one box with two partially filled tubes. No under-filled tubes were found in the other 5 boxes. DISCUSSIONThe blood to additive ratio is crucial for BCT. A sudden rise in clot errors should trigger a thorough investigation to identify the cause. MAIN LESSONLaboratories should regularly monitor and evaluate the percentage of clotted samples as a quality indicator at scheduled time points. A problem with the volume of additive in citrate tubes should be considered as a possible cause.
Bibliography:ObjectType-Case Study-2
content type line 59
SourceType-Reports-1
ObjectType-Report-1
ISSN:1846-7482
DOI:10.11613/BM.2019.021002