내시경기기 소독에 있어 과초산계 소독제(엔도파 Ⓡ )의 효과

Background/Aims: We aimed to investigate the efficacy of peracetic acid (EndoPA ® ; Firson Co., Ltd., Cheonan, Korea) in disinfecting endoscopes. Methods: We prospectively investigated the gastroscopes (Part I) utilized in 100 gastroscopic examinations and colonoscopes (Part II) utilized in 30 colon...

Full description

Saved in:
Bibliographic Details
Published in:The Korean journal of gastroenterology Vol. 71; no. 6; pp. 319 - 323
Main Authors: 이지민, Ji Min Lee, 이강문, Kang-moon Lee, 김대범, Dae Bum Kim, 고세은, Se Eun Go, 고성우, Sungwoo Ko, 강윤구, Yoongoo Kang, 홍소림, Solim Hong
Format: Journal Article
Language:Korean
Published: 대한소화기학회 30-06-2018
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Background/Aims: We aimed to investigate the efficacy of peracetic acid (EndoPA ® ; Firson Co., Ltd., Cheonan, Korea) in disinfecting endoscopes. Methods: We prospectively investigated the gastroscopes (Part I) utilized in 100 gastroscopic examinations and colonoscopes (Part II) utilized in 30 colonoscopic examinations after disinfecting them with 0.2% peracetic acid (EndoPA ® ; Firson Co., Ltd.). These instruments had been collected consecutively throughout the study period. We reprocessed and disinfected the endoscopes according to the guidelines for cleaning and disinfecting gastrointestinal endoscopes laid down by the Korean Society of Gastrointestinal Endoscopy in 2017. Three culture samples were obtained from each examination, based on different sampling methods. The primary outcome was a positive culture rate. Results: In Part I of our study, two of 300 samples were positive. The culture positive rate after disinfection was 0.7% (2/300). The culture positive rate was not significantly different based on the exposure time to EndoPA ® or the age of the scopes (p=0.7 or 0.2, respectively). In Part II of our study, all samples (n=90) were negative. Conclusions: We conclude that 0.2% peracetic acid (EndoPA ® ) appears to be a good disinfectant for both gastroscopes and colonoscopes. (Korean J Gastroenterol 2018;71:319-323)
Bibliography:Korean Society of Gastroenterology
ISSN:1598-9992