Reliability of the estimate of post-voiding bladder volume with ultrasound. Comparison of 2 ultrasonography methods
To determine the reliability of the postvoiding volume measurement of two US methods. A comparative study of transabdominal and portable ultrasound (Bladderscan model) in the selected and non-selected modes was conducted in 96 patients (54 males, 42 females; mean age 59.92 yrs). Conventional US had...
Saved in:
Published in: | Archivos españoles de urología Vol. 49; no. 1; p. 35 |
---|---|
Main Authors: | , , , , , , |
Format: | Journal Article |
Language: | Spanish |
Published: |
Spain
01-01-1996
|
Subjects: | |
Online Access: | Get more information |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To determine the reliability of the postvoiding volume measurement of two US methods.
A comparative study of transabdominal and portable ultrasound (Bladderscan model) in the selected and non-selected modes was conducted in 96 patients (54 males, 42 females; mean age 59.92 yrs).
Conventional US had the lowest failure rate (26.99%). The failure rate was 41.66% for the selected mode and 57.77% for the non-selected mode. The main failure of transabdominal ultrasound was a defect failure. The non-selected mode had a defect and excess failure. The selected mode had a defect failure. There was and inverse relationship between the residual volume and the failure rate of the non-selected mode of the Bladderscan model. The failure rate was higher for males vs females in the non-selected mode. The possible sources of failure are analyzed.
Conventional transabdominal ultrasound is currently the most reliable ultrasound method. |
---|---|