Impact of position of light mesh endoprosthesis with anisotropic structure for the efficiency of anterior abdominal wall reconstruction

Mechanical testing of light mesh endoprosthesis Ultrapro for tension and bending resistance showed strong anisotropy of the material along and across the loop columns. Ultrapro endoprosthesis was placed parallel and perpendicularly to the midline during plastic repair of a hernial defect in rats. Si...

Full description

Saved in:
Bibliographic Details
Published in:Bulletin of experimental biology and medicine Vol. 149; no. 6; pp. 779 - 783
Main Authors: Anurov, M V, Titkova, S M, Oettinger, A P
Format: Journal Article
Language:English
Published: United States Springer 01-11-2010
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Mechanical testing of light mesh endoprosthesis Ultrapro for tension and bending resistance showed strong anisotropy of the material along and across the loop columns. Ultrapro endoprosthesis was placed parallel and perpendicularly to the midline during plastic repair of a hernial defect in rats. Six months after surgery, the development of hernias and violation of endoprosthesis structure were observed in the group with longitudinal orientation of the endoprosthesis. Transverse orientation of the endoprosthesis led to fortification of the defect site. However, the meshes were in some cases misshapen and formed folds across the loop columns, because of insufficient bending rigidity and elasticity of the material.
Bibliography:ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Correction/Retraction-1
ObjectType-Feature-3
content type line 23
ISSN:0007-4888
1573-8221
DOI:10.1007/s10517-010-1051-8