Exit of catheter lock solutions from double lumen acute haemodialysis catheters—an in vitro study

Background. Double lumen dialysis catheters are routinely heparin or citrate ‘locked’ to maintain patency. Heparin lock-related bleeding episodes and antibiotic lock-related toxicity have been reported. The aim of this study is to quantify the amount of leak during ‘lock’ procedures and to compare l...

Full description

Saved in:
Bibliographic Details
Published in:Nephrology, dialysis, transplantation Vol. 22; no. 12; pp. 3533 - 3537
Main Authors: Sungur, Murat, Eryuksel, Emel, Yavas, Sinan, Bihorac, Azra, Layon, A. Joseph, Caruso, Lawrence
Format: Journal Article
Language:English
Published: Oxford Oxford University Press 01-12-2007
Oxford Publishing Limited (England)
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Background. Double lumen dialysis catheters are routinely heparin or citrate ‘locked’ to maintain patency. Heparin lock-related bleeding episodes and antibiotic lock-related toxicity have been reported. The aim of this study is to quantify the amount of leak during ‘lock’ procedures and to compare leakage for different double lumen dialysis catheters. Methods. In an experimental, in vitro study at a University research laboratory, five different double lumen dialysis catheters were tested using three different lock volumes. Results. Using the catheter flush volume, leak ratios for Flexxicon II 15 cm and 20 cm catheters were greater than that seen in the Arrow 16 cm catheter (P < 0.05). Using 20% less than the catheter flush volume, the Flexxicon II™ 20 cm catheter had greater leak than the Duo-flow 15 and 20 cm catheters and Arrow 16 cm catheter (P < 0.05). The Flexxicon II™ 15 cm catheter had greater leak than the Duo-flow 15 cm and Duo-flow 20 cm catheters with 20% less locking volume (P < 0.05). Using 20% greater than the catheter flush volume, the Duo-flow 20 cm catheter had significantly less leak ratio than the Flexxicon II™ 20 cm catheter (P < 0.05). There were no other significant differences in leak ratios between the catheters. Conclusion. All double lumen dialysis catheters we tested have a substantial amount of leak even when the catheter ‘lock’ volumes were used, and leak ratio increases significantly with 20% overfill. There is a leak even when using 20% less ‘lock’ volume. The amount of leak can be clinically important and may explain the reports of bleeding episodes after heparin lock and antibiotic toxicity after antibiotic and anticoagulant combination lock. Some devices have lower leak ratios than others, likely related to catheter design.
Bibliography:ark:/67375/HXZ-SZN2P3ZL-L
A part of this study was presented in the 10th Annual Conference on Continuous Renal Replacement Therapies, Dan Diego, California, 10–12 March 2005.
istex:DCF96D3987D58FB1C5F5271EAF02A4801D0E3E53
ISSN:0931-0509
1460-2385
DOI:10.1093/ndt/gfm452