A new radiofrequency thermal balloon catheter for pulmonary vein isolation

We sought to evaluate whether porcine pulmonary vein (PV) isolation (PVI) can be produced by ablation using our novel radiofrequency (RF) thermal balloon catheter (RBC). It has been proposed that PVI can prevent focal atrial fibrillation (AF) originating in or close to the PV. The RBC is composed of...

Full description

Saved in:
Bibliographic Details
Published in:Journal of the American College of Cardiology Vol. 38; no. 7; pp. 2079 - 2086
Main Authors: Tanaka, Kazushi, Satake, Shutaro, Saito, Shigeru, Takahashi, Saeko, Hiroe, Yoshitaka, Miyashita, Yusuke, Tanaka, Shinji, Tanaka, Michio, Watanabe, Yoshio
Format: Journal Article
Language:English
Published: New York, NY Elsevier Inc 01-12-2001
Elsevier Science
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We sought to evaluate whether porcine pulmonary vein (PV) isolation (PVI) can be produced by ablation using our novel radiofrequency (RF) thermal balloon catheter (RBC). It has been proposed that PVI can prevent focal atrial fibrillation (AF) originating in or close to the PV. The RBC is composed of a 12F main shaft, a 4F inner tube and a balloon. Inside the balloon, there is a unipolar coil electrode with a thermocouple sensor mounted along the tube, the former to deliver RF energy (13.56 MHz) and the latter to monitor the temperature. After the presence of a PV potential was confirmed, the RBC was safely inserted into the left atrium (LA) by the trans-septal approach. Once the balloon was inflated and optimally wedged at the junction between the PV and LA, RF energy was applied for 5 min. Radiofrequency catheter ablation (RFA) was repeated up to three times, until elimination of the PV potential or dissociation between the LA and PV was observed. Finally, each heart was examined histologically. In 18 PVs that had PV potentials, PVI was performed, resulting in success in 15 (success rate 83%, 95% confidence interval [CI] 58.0% to 96.3%; failure rate 17%, 95% CI 3.7% to 42.0%). After successful PVI, the PV potentials completely disappeared and the histologic examination revealed circumferential, transmural necrosis around the PV trunks. No major complications, such as PV stenosis or macroscopic thrombosis, were observed. The RBC was useful for PVI.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0735-1097
1558-3597
DOI:10.1016/S0735-1097(01)01666-7