Reusable Laryngeal Mask Airways can be used more than 40 times

Abstract Study Objective To compare the mechanical properties of classic, reusable Laryngeal Mask Airways (LMAs) that have been used more than 100 times with one unused LMA. Design Laboratory testing of devices used clinically. Setting Metropolitan university hospital. Measurements Cuff and tube spe...

Full description

Saved in:
Bibliographic Details
Published in:Journal of clinical anesthesia Vol. 20; no. 2; pp. 109 - 115
Main Authors: Goodman, Evan J., MD, Christenson, Elizabeth, PhD, Douglas, Allen M., CRNA, Ziegler, E. Jane, CRNA, Lewis, Bradley R., MD
Format: Journal Article
Language:English
Published: New York, NY Elsevier Inc 01-03-2008
Elsevier Science
Elsevier Limited
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Abstract Study Objective To compare the mechanical properties of classic, reusable Laryngeal Mask Airways (LMAs) that have been used more than 100 times with one unused LMA. Design Laboratory testing of devices used clinically. Setting Metropolitan university hospital. Measurements Cuff and tube specimens from LMAs that had been used at least 100 times were tested for elongation, tensile strength, stiffness, and tear strength using standard American Society of Testing and Materials protocols. Samples from an unused LMA were analyzed in the same manner for comparison. Main Results Tensile strength of the cuff samples was found to increase by approximately 25%. There was an increase in cuff stiffness and decreased tear strength similar to the manufacturer-reported trends. Cuff elongation decreased by 30%. Results for the LMA tube differed depending on whether the samples were taken in the machine direction or transverse direction. Tensile strength decreased by 30% in machine direction and, on average, very little in transverse direction. Conclusions The material in reusable classic LMAs does not lose its strength after 100 uses to the extent that its manufacturer claims. At least 100 uses may be considered safe for these devices.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0952-8180
1873-4529
DOI:10.1016/j.jclinane.2007.09.010