In vitro study of magnetic resonance imaging artefacts of six supraglottic airway devices

Summary We investigated the artefacts created during magnetic resonance imaging by five different laryngeal mask airways: the Classic (cLMA™); the LMA ProSeal™; the LMA Unique™; the Ambu® Disposable Laryngeal Mask; the LMA Supreme™; and one other supraglottic airway device, the i‐gel supraglottic ai...

Full description

Saved in:
Bibliographic Details
Published in:Anaesthesia Vol. 65; no. 6; pp. 569 - 572
Main Authors: Zaballos, M., Bastida, E., Del Castillo, T., De Villoria, J. G., Jiménez, C.
Format: Journal Article
Language:English
Published: Oxford, UK Blackwell Publishing Ltd 01-06-2010
Wiley-Blackwell
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Summary We investigated the artefacts created during magnetic resonance imaging by five different laryngeal mask airways: the Classic (cLMA™); the LMA ProSeal™; the LMA Unique™; the Ambu® Disposable Laryngeal Mask; the LMA Supreme™; and one other supraglottic airway device, the i‐gel supraglottic airway. The devices were placed on top of and inside a phantom simulator to resemble the position in vivo. The artefacts with the cLMA, Unique and Supreme were similar and related to ferromagnetic material in the pilot balloon valve. Artefacts were more prominent with the ProSeal. There were no artefacts with the Ambu Disposable Laryngeal Mask or the i‐gel.