Detection of expiratory flow limitation in COPD using the forced oscillation technique

Expiratory flow limitation (EFL) during tidal breathing is a major determinant of dynamic hyperinflation and exercise limitation in chronic obstructive pulmonary disease (COPD). Current methods of detecting this are either invasive or unsuited to following changes breath-by-breath. It was hypothesis...

Full description

Saved in:
Bibliographic Details
Published in:The European respiratory journal Vol. 23; no. 2; pp. 232 - 240
Main Authors: Dellaca, R.L, Santus, P, Aliverti, A, Stevenson, N, Centanni, S, Macklem, P.T, Pedotti, A, Calverley, P.M.A
Format: Journal Article
Language:English
Published: Leeds Eur Respiratory Soc 01-02-2004
Maney
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Expiratory flow limitation (EFL) during tidal breathing is a major determinant of dynamic hyperinflation and exercise limitation in chronic obstructive pulmonary disease (COPD). Current methods of detecting this are either invasive or unsuited to following changes breath-by-breath. It was hypothesised that tidal flow limitation would substantially reduce the total respiratory system reactance (Xrs) during expiration, and that this reduction could be used to reliably detect if EFL was present. To test this, 5-Hz forced oscillations were applied at the mouth in seven healthy subjects and 15 COPD patients (mean +/- sD forced expiratory volume in one second was 36.8 +/- 11.5% predicted) during quiet breathing. COPD breaths were analysed (n=206) and classified as flow-limited if flow decreased as alveolar pressure increased, indeterminate if flow decreased at constant alveolar pressure, or nonflow-limited. Of these, 85 breaths were flow-limited, 80 were not and 41 were indeterminate. Among other indices, mean inspiratory minus mean expiratory Xrs (deltaXrs) and minimum expiratory Xrs (Xexp,min) identified flow-limited breaths with 100% specificity and sensitivity using a threshold between 2.53-3.12 cmH2O x s x L(-1) (deltaXrs) and -7.38- -6.76 cmH2O x s x L(-1) (Xexp,min) representing 6.0% and 3.9% of the total range of values respectively. No flow-limited breaths were seen in the normal subjects by either method. Within-breath respiratory system reactance provides an accurate, reliable and noninvasive technique to detect expiratory flow limitation in patients with chronic obstructive pulmonary disease.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0903-1936
1399-3003
DOI:10.1183/09031936.04.00046804