A compact high resolution flat panel PET detector based on the new 4-side buttable MPPC for biomedical applications
Compact high-resolution panel detectors using virtual pinhole (VP) PET geometry can be inserted into existing clinical or pre-clinical PET systems to improve regional spatial resolution and sensitivity. Here we describe a compact panel PET detector built using the new Though Silicon Via (TSV) multi-...
Saved in:
Published in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Vol. 794; pp. 151 - 159 |
---|---|
Main Authors: | , , , , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
Netherlands
Elsevier B.V
11-09-2015
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Compact high-resolution panel detectors using virtual pinhole (VP) PET geometry can be inserted into existing clinical or pre-clinical PET systems to improve regional spatial resolution and sensitivity. Here we describe a compact panel PET detector built using the new Though Silicon Via (TSV) multi-pixel photon counters (MPPC) detector. This insert provides high spatial resolution and good timing performance for multiple bio-medical applications. Because the TSV MPPC design eliminates wire bonding and has a package dimension which is very close to the MPPC׳s active area, it is 4-side buttable. The custom designed MPPC array (based on Hamamatsu S12641-PA-50(x)) used in the prototype is composed of 4×4 TSV-MPPC cells with a 4.46mm pitch in both directions. The detector module has 16×16 lutetium yttrium oxyorthosilicate (LYSO) crystal array, with each crystal measuring 0.92×0.92×3mm3 with 1.0mm pitch. The outer diameter of the detector block is 16.8×16.8mm2. Thirty-two such blocks will be arranged in a 4×8 array with 1mm gaps to form a panel detector with detection area around 7cm×14cm in the full-size detector. The flood histogram acquired with 68Ge source showed excellent crystal separation capability with all 256 crystals clearly resolved. The detector module׳s mean, standard deviation, minimum (best) and maximum (worst) energy resolution were 10.19%, ±0.68%, 8.36% and 13.45% FWHM, respectively. The measured coincidence time resolution between the block detector and a fast reference detector (around 200 ps single photon timing resolution) was 0.95ns. When tested with Siemens Cardinal electronics the performance of the detector blocks remain consistent. These results demonstrate that the TSV-MPPC is a promising photon sensor for use in a flat panel PET insert composed of many high resolution compact detector modules. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0168-9002 1872-9576 |
DOI: | 10.1016/j.nima.2015.04.071 |