Formation and Evaluation of Complete Blood Count Proficiency Testing Program
: The haematology external quality assessment (EQA) scheme is the most commonly used service of quality assurance. The provision of complete blood count (CBC) materials must meet the quality requirements at a reasonable cost. These requirements are the most significant challenges for EQA organisers...
Saved in:
Published in: | Hematology reports Vol. 14; no. 2; pp. 73 - 84 |
---|---|
Main Authors: | , , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
Switzerland
MDPI AG
25-03-2022
MDPI |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | : The haematology external quality assessment (EQA) scheme is the most commonly used service of quality assurance. The provision of complete blood count (CBC) materials must meet the quality requirements at a reasonable cost. These requirements are the most significant challenges for EQA organisers in Vietnam. This study's objective was to evaluate the homogeneity, long-term stability, and peer-group performance of 10-parameter stabilised CBC EQA samples.
: The CBC EQA material was prepared using the following steps, including (1) adjusting levels of stabilised erythrocyte, leukocyte, and platelet samples, (2) mixing those cells into batches at three levels, and (3) dispensing and storing them at 2-6 °C. A set of 10 and 30 specimens were randomly chosen from each batch to study the homogeneity and long-term stability following ISO 13528:2015. In total, 166 samples at two levels were randomly distributed to 40 participants, which reported 83 automatic cell counters among six automated analyser models in the CBC EQA program.
: The 10-parameter stabilised CBC EQA materials at three levels became homogeneous and stable in 12 weeks when preserved at 2-6 °C. Meanwhile, for five parameters (RBC, Hb, MCH, MCV, and MPV), this process was prolonged for up to 16 weeks in stock condition. In terms of peer-group performance, the CV (%) values increased at the low concentration for almost all parameters, especially in platelet counts.
: The stabilised CBC EQA samples prepared using the partial fixation method with aldehyde and gutaraldehyde in this study meet the ISO 13528:2015 requirements of homogeneity and long-term stability for the CBC EQA scheme. Analytical performance evaluation should categorise participant methods into peer groups. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 2038-8322 2038-8330 2038-8330 |
DOI: | 10.3390/hematolrep14020012 |