Clinical Performance of the St. Jude Medical Riata Defibrillation Lead in a Large Patient Population

Clinical Performance of the St. Jude Medical Riata Defibrillation Lead in a Large Patient Population. Objective: The purpose of this large multicenter study was to evaluate the long‐term reliability of an implantable cardioverter defibrillator (ICD) lead to determine the incidence of adverse events...

Full description

Saved in:
Bibliographic Details
Published in:Journal of cardiovascular electrophysiology Vol. 21; no. 5; pp. 551 - 556
Main Authors: PORTERFIELD, JAMES G., PORTERFIELD, LINDA M., KUCK, KARL H., CORBISIERO, RAFFAELE, GREENBERG, STEVEN M., HINDRICKS, GERHARD, WAZNI, OUSSAMA, BEAU, SCOTT L., HERRE, JOHN M.
Format: Journal Article
Language:English
Published: Malden, USA Blackwell Publishing Inc 01-05-2010
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Clinical Performance of the St. Jude Medical Riata Defibrillation Lead in a Large Patient Population. Objective: The purpose of this large multicenter study was to evaluate the long‐term reliability of an implantable cardioverter defibrillator (ICD) lead to determine the incidence of adverse events (AEs). Background: A recent concern has been the performance of cardiac defibrillator leads. There have been conflicting reports regarding the rate of lead perforation and other AEs. Methods: Medical records from patients implanted from 6‐1‐2001 to 11‐27‐2007 with the St. Jude Medical Riata family of RV leads at 23 US (N = 12,969) and 5 German (N = 2,418) centers were reviewed for chronic lead‐related AEs. These included perforation, dislodgment, conductor fracture and insulation damage. The mean follow‐up period was 18.0 months. AEs were defined as those that required Riata lead revision, extraction, or replacement. Results: The incidence of lead AEs was <1% for each AE type. Perforation occurred in 0.38%, dislodgement in 0.93%, conductor fracture in 0.18%, and insulation damage in 0.21% of patients studied. Conclusions: During the follow‐up of the 15,387 patients with Riata leads, the incidence of AEs which included perforation, dislodgement, conductor fraction and insulation damage was low and within the range of what is considered clinically acceptable. (J Cardiovasc Electrophysiol, Vol. 21, pp. 551‐556, May 2010)
Bibliography:ark:/67375/WNG-5GTFK67T-2
istex:604295777101E2F754CEE03E017BABFF25AFA8B4
ArticleID:JCE1651
Disclosures
J. Porterfield: St. Jude Medical (Grants/Research Support); L. Porterfield: St. Jude Medical (Grants/Research Support); KH Kuck: St. Jude Medical (Consultancies/Honorariums), Stereotaxis (Consultancies/Honorariums) Biosense Webster (Research Agreements), Cryocath (Research Agreements), EHRA (Fellowship Support); R. Corbisiero: St. Jude Medical (Consultancies/Honorariums, Research Agreements); S. Greenberg: St. Jude Medical (Consultancies/Honorariums, Research Agreements, Stock Ownership, Corporate Appointments, Speaker, Advisory Board), Medtronic (Speaker, Advisory Board), BSX (Speaker); G. Hindricks: St. Jude Medical (Research Support, Speaker, Advisory Board); O. Wazni: St. Jude Medical (Research Support, Consultancies/Honorariums), BSX; (Consultancies/Honorariums), Medtronic (Consultancies/Honorariums), Transoma (Research Agreements), eCardio (Corporate Appointments, Advisory Boards); S. Beau: St. Jude Medical (Consultancies/Honorariums); J. Herre: St. Jude Medical (Consultancies/Honorariums, Research Agreements), Medtronic (Consultancies/Honorariums, Research Agreements), Boston Scientific (Consultancies/Honorariums, Research Agreements), Medtronic (Corporate Appointments), LifeNet Health (Chairman of the Board).
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1045-3873
1540-8167
DOI:10.1111/j.1540-8167.2009.01651.x