Development of a Type-N Coaxial Microcalorimeter for RF and Microwave Power Standards at KRISS

This paper introduces a newly constructed type-N coaxial microcalorimeter that will serve as a primary power standard in Korea Research Institute of Standards and Science. The detailed design scheme of the microcalorimeter is presented, and its characteristics are evaluated. The performance of an ad...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on instrumentation and measurement Vol. 64; no. 6; pp. 1520 - 1526
Main Authors: Jae-Yong Kwon, Tae-Weon Kang, Jeong-Hwan Kim, No-Weon Kang
Format: Journal Article
Language:English
Published: IEEE 01-06-2015
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:This paper introduces a newly constructed type-N coaxial microcalorimeter that will serve as a primary power standard in Korea Research Institute of Standards and Science. The detailed design scheme of the microcalorimeter is presented, and its characteristics are evaluated. The performance of an adiabatic line was simulated with a finite-element-method-based tool and predicted. The measured insertion loss of the assembled adiabatic line was less than 0.15 dB. Frequency and power sweep measurements were performed on a reference standard to find a substituted power level that minimized the dual-element error. We also propose evaluation standards that can be used for the measurement of radio frequency loss. The correction factor for the effective efficiency of the reference standard was experimentally evaluated. We also present an uncertainty evaluation process and the corrected effective efficiency of a type-N coaxial reference standard.
ISSN:0018-9456
1557-9662
DOI:10.1109/TIM.2015.2408802