Evaluation of a long-range transmitter for use with a magnetic tracking device in motion analysis
The suitability of a long-range transmitter was evaluated for use with a Polhemus Fastrak ® magnetic tracking device in kinesiologic studies. Performance was judged by comparing positional and rotational accuracy to a standard transmitter. Data were obtained at distances of up to 2.7 and 5.0 m for t...
Saved in:
Published in: | Journal of biomechanics Vol. 31; no. 10; pp. 957 - 961 |
---|---|
Main Authors: | , , |
Format: | Journal Article |
Language: | English |
Published: |
United States
Elsevier Ltd
01-10-1998
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The suitability of a long-range transmitter was evaluated for use with a Polhemus Fastrak
® magnetic tracking device in kinesiologic studies. Performance was judged by comparing positional and rotational accuracy to a standard transmitter. Data were obtained at distances of up to 2.7 and 5.0
m for the standard and long-range transmitters, respectively. Use of the long-range transmitter improved system performance allowing reproducible measurements at a greater distance. However, it is necessary to calibrate the system in each new test environment as there can be significant distortion of the magnetic field. |
---|---|
Bibliography: | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 ObjectType-Article-1 ObjectType-Feature-2 |
ISSN: | 0021-9290 1873-2380 |
DOI: | 10.1016/S0021-9290(98)00089-X |