Measuring anterior chamber depth with Orbscan Topography System
To assess the reliability of the Orbscan Topography System in measuring anterior chamber depth (ACD). Department of Ophthalmology, Ruprecht-Karls-University of Heidelberg, Heidelberg, Germany. The Orbscan Topography System is a 3-D, scanning slit beam system for analyzing corneal surfaces as well as...
Saved in:
Published in: | Journal of cataract and refractive surgery Vol. 23; no. 9; p. 1351 |
---|---|
Main Authors: | , , , |
Format: | Journal Article |
Language: | English |
Published: |
United States
01-11-1997
|
Subjects: | |
Online Access: | Get more information |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To assess the reliability of the Orbscan Topography System in measuring anterior chamber depth (ACD).
Department of Ophthalmology, Ruprecht-Karls-University of Heidelberg, Heidelberg, Germany.
The Orbscan Topography System is a 3-D, scanning slit beam system for analyzing corneal surfaces as well as structures of the anterior segment, e.g., iris, lens. Reflected surface data points are measured in the x, y, and z axes, creating color-coded true surface topography maps. We measured ACD prior to cataract surgery in 56 eyes of 37 patients (mean age 70.4 years + 13.2 [SD]) using the Orbscan system and, as a control, the Grieshaber Biometric System (ultrasound) and the Jäger ACD slitlamp measurement system.
Mean ACD values measured with the Orbscan system were 3.23 +/- 0.55 mm; ultrasound measurements were 3.27 +/- 0.54 mm and Jäger measurements, 3.10 +/- 0.55 mm. The average difference between Orbscan and ultrasound values was 0.04 +/- 0.15 mm. The correlation coefficient between Orbscan and ultrasound was 0.96 (P < .00001) and between Orbscan and Jäger, 0.899 (P < .0001).
The ACD measurements by the Orbscan system were equivalent to common reference measurements. |
---|---|