Estimation of Arterial Carbon Dioxide Based on End-Tidal Gas Pressure and Oxygen Saturation
Arterial blood gas (ABG) analysis is the traditional method for measuring the partial pressure of carbon dioxide. In mechanically ventilated patients a continuous noninvasive monitoring of would obviously be attractive. In the current study, we present a novel formula for noninvasive estimation of a...
Saved in:
Published in: | Journal of clinical medicine Vol. 7; no. 9; p. 290 |
---|---|
Main Authors: | , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
Switzerland
MDPI
19-09-2018
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Arterial blood gas (ABG) analysis is the traditional method for measuring the partial pressure of carbon dioxide. In mechanically ventilated patients a continuous noninvasive monitoring of
would obviously be attractive. In the current study, we present a novel formula for noninvasive estimation of arterial carbon dioxide. Eighty-one datasets were collected from 19 anesthetized and mechanically ventilated pigs. Eleven animals were mechanically ventilated without interventions. In the remaining eight pigs the partial pressure of carbon dioxide was manipulated. The new formula (Formula 1) is PaCO₂ = PETCO₂ + k(PETO₂ - PaO₂) where PaO₂ was calculated from the oxygen saturation. We tested the agreements of this novel formula and compared it to a traditional method using the baseline PaCO₂ - ETCO₂ gap added to subsequently measured, end-tidal carbon dioxide levels (Formula 2). The mean difference between PaCO₂ and calculated carbon dioxide (Formula 1) was 0.16 kPa (±SE 1.17). The mean difference between PaCO₂ and carbon dioxide with Formula 2 was 0.66 kPa (±SE 0.18). With a mixed linear model excluding cases with cardiorespiratory collapse, there was a significant difference between formulae (
< 0.001), as well as significant interaction between formulae and time (
< 0.001). In this preliminary animal study, this novel formula appears to have a reasonable agreement with PaCO₂ values measured with ABG analysis, but needs further validation in human patients. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 2077-0383 2077-0383 |
DOI: | 10.3390/jcm7090290 |