A novel procedure of secondary voice prosthesis insertion from the inside out: The modified Fukuhara method

To present a new procedure for inserting a secondary voice prosthesis from the inside to the outside, which improves on the method previously reported by Fukuhara et al. A flexible nasopharyngoscope was used to puncture pharynx (or transplanted jejunum) from the inside to the outside. In this method...

Full description

Saved in:
Bibliographic Details
Published in:Auris, nasus, larynx Vol. 49; no. 4; pp. 658 - 662
Main Authors: Hanai, Nobuhiro, Beppu, Shintaro, Nishikawa, Daisuke, Terada, Hoshino, Sawabe, Michi
Format: Journal Article
Language:English
Published: Netherlands Elsevier B.V 01-08-2022
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:To present a new procedure for inserting a secondary voice prosthesis from the inside to the outside, which improves on the method previously reported by Fukuhara et al. A flexible nasopharyngoscope was used to puncture pharynx (or transplanted jejunum) from the inside to the outside. In this method, it was possible to use the PROVOX® VegaTM Puncture Set as it is used for the placement of the voice prosthesis. We were able to place the PROVOX® VegaTM in all cases we experienced. Most of the cases had a history of radiation therapy. The time required for surgery ranged from 11–59 minutes (mean: 29 minutes) and there was no measurable amount of bleeding. This new method using the PROVOX® VegaTM Puncture Set, which is designed for the original purpose of voice prosthesis implantation, was therefore found to be safe and effective.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0385-8146
1879-1476
DOI:10.1016/j.anl.2021.11.006