Improved Accuracy of Noninvasive Electron Beam Coronary Angiography
OBJECTIVES:We investigated the effect of electrocardiographic (ECG) triggering on the accuracy of coronary electron-beam angiography (EBA) as compared with invasive angiography. METHODS:One hundred thirty-three patients with suspected coronary disease were studied with intravenous coronary EBA and c...
Saved in:
Published in: | Investigative radiology Vol. 39; no. 2; pp. 73 - 79 |
---|---|
Main Authors: | , , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
United States
Lippincott Williams & Wilkins, Inc
01-02-2004
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | OBJECTIVES:We investigated the effect of electrocardiographic (ECG) triggering on the accuracy of coronary electron-beam angiography (EBA) as compared with invasive angiography.
METHODS:One hundred thirty-three patients with suspected coronary disease were studied with intravenous coronary EBA and conventional coronary angiography. Patients were divided into 2 groups based upon ECG triggering on the EBA study. Patients were divided into 2 groups based upon different ECG triggering used80% R-R interval trigger method (group 1, n = 53) and end-systolic triggering (group 2, n = 80). End-systolic ECG triggering, which started at the end of the T wave in each study, was based on baseline heart rate.
RESULTS:Overall sensitivity to detect a ≥50% luminal stenosis was 69% in group 1 and 91% in group 2 (P = 0.002); specificity was 82% and 94% in group 1 and group 2, respectively (P < 0.001). Using newer triggering techniques (group 2) with EBA, the sensitivity, specificity, and accuracy for patients with disease of the left main coronary artery or 3 vessel disease was 100%, 94%, and 98%, respectively. Nonassessability of coronary segments on 3D-EBA images was reduced from 35% in group 1 to 9% in group 2 patients (P < 0.001). The number of motion-free coronary images increased from 67% to 95% from group 1 to group 2 (P < 0.0001).
CONCLUSION:End-systolic ECG triggering improves accuracy, image quality, and assessability of segments of coronary EBA for the detection of angiographic coronary artery disease. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0020-9996 1536-0210 |
DOI: | 10.1097/01.rli.0000105330.17743.c5 |