Assessment of a circular powered stapler for creation of anastomosis in left-sided colorectal surgery: A prospective cohort study
Circular staplers perform a critical function for creation of anastomoses in colorectal surgeries. Powered stapling systems allow for reduced force required by surgeons to fire the device and may provide advantages for creating a secure anastomosis. The objective of this study was to evaluate the cl...
Saved in:
Published in: | International journal of surgery (London, England) Vol. 84; pp. 140 - 146 |
---|---|
Main Authors: | , , , , , , , , , , , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
England
Elsevier Ltd
01-12-2020
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Circular staplers perform a critical function for creation of anastomoses in colorectal surgeries. Powered stapling systems allow for reduced force required by surgeons to fire the device and may provide advantages for creating a secure anastomosis. The objective of this study was to evaluate the clinical performance of a novel circular powered stapler in a post-market setting, during left-sided colectomy procedures.
Consecutive subjects underwent left-sided colorectal resections that included anastomosis performed with the ECHELON CIRCULAR™ Powered Stapler (ECP). The primary endpoint was the frequency in which a stapler performance issue was observed. Secondary endpoints included evaluation of ease of use of the device via a surgeon satisfaction questionnaire, and monitoring/recording of procedure-related adverse events (AEs).
A total of 168 anastomoses were performed with the ECP. Surgical approaches included robotic-assisted (n = 74, 44.0%), laparoscopic (n = 71, 42.3%), open (n = 20, 11.9%), and hand-assisted minimally invasive (n = 3, 1.8%) procedures. There were 22 occurrences of device performance issues in 20 (11.9%) subjects during surgery. No positive intraoperative leak tests were observed, and only 1 issue was related to a procedure-related AE or surgical complication, which was an instance of incomplete surgical donut necessitating re-anastomosis. Postoperative anastomotic leaks were experienced in 4 (2.4%) subjects. Clavien-Dindo classification of all AEs indicated that 92.0% were Grades I or II. Participating surgeons rated the ECP as easier to use compared to previously used manual circular staplers in 85.7% of procedures.
The circular powered stapler exhibited few clinically relevant performance issues, an overall favorable safety profile, and ease of use for creation of left-sided colon anastomoses.
•Test device performance issues occurred in 20 (11.9%) subjects with only one issue requiring surgical revision.•There were zero (0%) intraoperative leaks and 4 (2.4%) postoperative anastomotic leaks among 168 procedures.•The circular powered stapler demonstrated ease of use as judged by investigators. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 1743-9191 1743-9159 |
DOI: | 10.1016/j.ijsu.2020.11.001 |