Application of 1,000 fps High-Speed Angiography to In-Vitro Hemodynamic Evaluation of Left Ventricular Assist Device Outflow Graft Configurations

Left ventricular assist device (LVAD)–induced hemodynamics are characterized by fast-moving flow with large variations in velocity, making quantitative assessments difficult with existing imaging methods. This study demonstrates the ability of 1,000 fps high-speed angiography (HSA) to quantify the e...

Full description

Saved in:
Bibliographic Details
Published in:ASAIO journal (1992) Vol. 69; no. 8; pp. 756 - 765
Main Authors: Shields, Allison, Setlur Nagesh, Swetadri Vasan, Rajagopal, Keshava, Bednarek, Daniel R., Rudin, Stephen, Chivukula, Venkat Keshav
Format: Journal Article
Language:English
Published: United States Lippincott Williams & Wilkins 01-08-2023
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Left ventricular assist device (LVAD)–induced hemodynamics are characterized by fast-moving flow with large variations in velocity, making quantitative assessments difficult with existing imaging methods. This study demonstrates the ability of 1,000 fps high-speed angiography (HSA) to quantify the effect of the surgical implantation angle of a LVAD outflow graft on the hemodynamics within the ascending aorta in vitro. High-speed angiography was performed on patient-derived, three-dimensional–printed optically opaque aortic models using a nonsoluble contrast media, ethiodol, as a flow tracer. Outflow graft configuration angles of 45° and 90° with respect to the central aortic axis were considered. Projected velocity distributions were calculated from the high-speed experimental sequences using two methodsa physics-based optical flow algorithm and tracking of radio-opaque particles. Particle trajectories were also used to evaluate accumulated shear stress. Results were then compared with computational fluid dynamics (CFD) simulations to confirm the results of the high-speed imaging method. Flow patterns derived from HSA coincided with the impingement regions and recirculation zones formed in the aortic root as seen in the CFD for both graft configurations. Compared with the 45° graft, the 90° configuration resulted in 81% higher two-dimensional–projected velocities (over 100 cm/s) along the contralateral wall of the aorta. Both graft configurations suggest elevated accumulated shear stresses along individual trajectories. Compared with CFD simulations, HSA successfully characterized the fast-moving flow and hemodynamics in each LVAD graft configuration in vitro, demonstrating the potential utility of this technology as a quantitative imaging modality.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1058-2916
1538-943X
1538-943X
DOI:10.1097/MAT.0000000000001948