Accuracy and Cost-Effectivenss of a Novel Method for Alpha Defensins Measurement in the Diagnosis of Periprosthetic Joint Infections
Two methods for detecting synovial fluids alpha defensins are available: the enzyme-linked immunosorbent assay and the lateral flow test. For both, the proper role and accuracy remain uncertain. The purpose of this study was to assess the accuracy of the matrix-assisted laser desorption/ionization t...
Saved in:
Published in: | The Journal of arthroplasty Vol. 36; no. 9; pp. 3275 - 3281 |
---|---|
Main Authors: | , , , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
Elsevier Inc
01-09-2021
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Two methods for detecting synovial fluids alpha defensins are available: the enzyme-linked immunosorbent assay and the lateral flow test. For both, the proper role and accuracy remain uncertain. The purpose of this study was to assess the accuracy of the matrix-assisted laser desorption/ionization time of flight mass spectrometry (MALDI-TOF MS) for alpha defensin detection in synovial fluids of patients with total knee arthroplasty/total hip arthroplasty failures. The hypothesis was that the alpha defensin measurement through MALDI-TOF MS assay could be a high sensitive and specific test for periprosthetic joint infections (PJI) diagnosis as compared with Musculoskeletal Infection Society (MSIS) criteria.
The study included 138 patients. The 2018 MSIS criteria were used to diagnose PJIs. Synovial fluids were assessed for routinely synovial fluid tests and alpha defensin measurement through MALDI-TOF MS. Sensitivity, specificity, overall diagnostic accuracy, positive and negative predictive values, receiver operator curves, and area under the curve were calculated.
As per the 2018 MSIS criteria, 59 PJIs (43%) and 79 aseptic failures (57%) were diagnosed. The MALDI-TOF MS assay showed an overall accuracy of 94.9%. The sensitivity was 93%, the specificity was 96%, the positive predictive value was 95%, and the negative predictive value was 95%. Receiver operator curves analysis demonstrates an area under the curve of 0.95 (P < .001).
The MALDI-TOF MS assay showed high sensitivity and specificity for alpha defensin detection in case of total knee arthroplasty/total hip arthroplasty failures. The advantages of the technology, such as the few milliliters of sample needed, the rapidity of obtaining results, and the cost-effectiveness of the procedure could make the MALDI-TOF MS alpha defensin assay a useful and widespread test in clinical practice. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0883-5403 1532-8406 |
DOI: | 10.1016/j.arth.2021.05.013 |