The liquid nitrogen fill level meter for the AGATA triple cluster detector
A novel liquid nitrogen fill level meter has been put into operation for the all-position dewar of the triple cluster detector of the Advanced GAmma Tracking Array. The new device is based on a capacitance measurement between a metallic cylindrical tube inside the dewar and the inner wall of the cry...
Saved in:
Published in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Vol. 640; no. 1; pp. 133 - 138 |
---|---|
Main Authors: | , , , , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
Elsevier B.V
01-06-2011
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A novel liquid nitrogen fill level meter has been put into operation for the all-position dewar of the triple cluster detector of the Advanced GAmma Tracking Array. The new device is based on a capacitance measurement between a metallic cylindrical tube inside the dewar and the inner wall of the cryostat. The fill level dependent capacitance is converted by a
C/
V-transducer into a DC voltage signal. Direct monitoring of the LN
2 level inside the detector dewar has been performed with several AGATA detectors at various inclinations and rotation angles of the detector axis. The time-dependent LN
2 consumption is an additional quantity used to survey the status of the cryostat. Supplementary results are the investigations of the LN
2 consumption and the heat loss of the detector during different modes of operation. |
---|---|
Bibliography: | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 |
ISSN: | 0168-9002 1872-9576 |
DOI: | 10.1016/j.nima.2011.02.080 |