Performance of Multisphere Spectrometry Systems
A TLD-based multisphere spectrometer has been constructed to measure neutron spectra under high fluence-rate conditions. The multisphere system uses paired TLD-600 and TLD-700 thermoluminescence detectors. Spectral measurements made with this system are compared with measurements made with a 6LiI(Eu...
Saved in:
Published in: | Radiation protection dosimetry Vol. 78; no. 4; pp. 263 - 272 |
---|---|
Main Authors: | , , , |
Format: | Journal Article |
Language: | English |
Published: |
Oxford
Oxford University Press
01-01-1998
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A TLD-based multisphere spectrometer has been constructed to measure neutron spectra under high fluence-rate conditions. The multisphere system uses paired TLD-600 and TLD-700 thermoluminescence detectors. Spectral measurements made with this system are compared with measurements made with a 6LiI(Eu) scintillator-based Bonner sphere spectrometer. The system was used to measure the neutron spectrum of the Georgia Tech Research Reactor Bio-Medical Facility under high neutron fluence rates. The scintillator-based system was used at low neutron and gamma fluence rates. A comparison of this unfolded spectrum is made with an MCNP code calculation performed using a simplified representation of the reactor and the beam facility. |
---|---|
Bibliography: | local:780263 PII:0144-8420 ark:/67375/HXZ-ZQ7S13SJ-K istex:61198B24DBB5334593EB03BD13E1769F4A207598 |
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/oxfordjournals.rpd.a032359 |