Use of two parallel oxygenators in a very large patient (2.76 m2) for an acute "A" dissecting aortic aneurysm repair

The very large patient (weight 142 kg, height 197 cm, body surface 2.76 m2) was referred to acute operation with dissecting type A ascending aortic aneurysm. The calculated blood flow was 6.63 l/min. To anticipate potential difficulties with perfusion and oxygenation two oxygenators connected in par...

Full description

Saved in:
Bibliographic Details
Published in:Acta medica Lékarskí fakulty Univerzity Karlovy v Hradci Králove Vol. 48; no. 2; pp. 95 - 98
Main Authors: Lonský, Vladimir, Mand'ák, Jiri, Kubícek, Jaroslav, Volt, Martin, Procházka, Egon, Dominik, Jan
Format: Journal Article
Language:English
Published: Czech Republic Karolinum Press 2005
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The very large patient (weight 142 kg, height 197 cm, body surface 2.76 m2) was referred to acute operation with dissecting type A ascending aortic aneurysm. The calculated blood flow was 6.63 l/min. To anticipate potential difficulties with perfusion and oxygenation two oxygenators connected in parallel were incorporated into the circuit. Bentall procedure with ACB to the RCA was performed. The perfusion was uneventful. Bypass time was 259 minutes, cross clamp time 141 minutes, circulatory arrest 7 minutes. The highest oxygenators gas flow was 2.6 l/min with maximum FiO2 0.42. The use of two in parallel connected oxygenators is a very effective, easy and safe method in such extreme perfusions, offering to the perfusionist a great reserve of oxygenator output.
Bibliography:ObjectType-Case Study-2
SourceType-Scholarly Journals-1
ObjectType-Feature-4
content type line 23
ObjectType-Report-1
ObjectType-Article-3
ISSN:1211-4286
1805-9694
DOI:10.14712/18059694.2018.39