Adjustable compensators for whole-brain irradiation

A compensator for use in irradiating the whole brain with parellel opposed lateral fields was made from a stack of thin tempered aluminum plates. This configuration gives a wide latitude of degrees of compensation and can accommodate a broad range of head sizes. Dosimetry studies indicate that overc...

Full description

Saved in:
Bibliographic Details
Published in:Radiology Vol. 130; no. 2; p. 529
Main Authors: Lerch, I A, Newall, J
Format: Journal Article
Language:English
Published: United States 01-02-1979
Subjects:
Online Access:Get more information
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A compensator for use in irradiating the whole brain with parellel opposed lateral fields was made from a stack of thin tempered aluminum plates. This configuration gives a wide latitude of degrees of compensation and can accommodate a broad range of head sizes. Dosimetry studies indicate that overcompensation may be desirable to reduce the dose to the mid-coronal plane of the skull and scalp without seriously affecting the dose uniformity in the treatment volume.