A double dc SQUID based magnetometer
We have developed a double dc SQUID based magnetometer with a direct read-out circuit and measured its signal and noise characteristics. The inductances of the input SQUID and SQUID-preamplifier were 120 and 250 pH. Thin-film flux transformers on the separate substrates had an inductance of the orde...
Saved in:
Published in: | IEEE transactions on applied superconductivity Vol. 5; no. 2; pp. 3241 - 3243 |
---|---|
Main Authors: | , , , , |
Format: | Journal Article |
Language: | English |
Published: |
IEEE
01-06-1995
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We have developed a double dc SQUID based magnetometer with a direct read-out circuit and measured its signal and noise characteristics. The inductances of the input SQUID and SQUID-preamplifier were 120 and 250 pH. Thin-film flux transformers on the separate substrates had an inductance of the order of 1 /spl mu/H each and a current sensitivity of 0.54 and 0.32 /spl mu/A//spl Phi//sub 0/ for the first and the second SQUIDs, respectively. The intrinsic energy resolution of the fabricated magnetometer for an optimal gain of 58 for the first stage was close to 1.6/spl times/10/sup -32/ J/Hz. The dynamic range and the slew rate were about 126 dB and 4/spl times/10/sup 4/ /spl Phi//sub 0//s.< > |
---|---|
Bibliography: | SourceType-Scholarly Journals-2 ObjectType-Feature-2 ObjectType-Conference Paper-1 content type line 23 SourceType-Conference Papers & Proceedings-1 ObjectType-Article-3 ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 |
ISSN: | 1051-8223 1558-2515 |
DOI: | 10.1109/77.403282 |