Detection of oxandrolone and its metabolite in urine by high-performance liquid chromatography-high-resolution mass spectrometry with atmospheric pressure chemical ionization and orbitrap detection after ceasing drug administration
It was shown that oxandrolone can be detected by high-performance liquid chromatography-high-resolution mass spectrometry with atmospheric pressure chemical ionization and orbitrap detection up to two weeks after ceasing its administration. The mass accuracy in the analysis of real samples was 2 ppm...
Saved in:
Published in: | Journal of analytical chemistry (New York, N.Y.) Vol. 64; no. 1; pp. 31 - 35 |
---|---|
Main Authors: | , , |
Format: | Journal Article |
Language: | English |
Published: |
Dordrecht
SP MAIK Nauka/Interperiodica
2009
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | It was shown that oxandrolone can be detected by high-performance liquid chromatography-high-resolution mass spectrometry with atmospheric pressure chemical ionization and orbitrap detection up to two weeks after ceasing its administration. The mass accuracy in the analysis of real samples was 2 ppm. The detection limits for oxandrolone and its metabolite were 3 and 1 pg/mL, respectively (signal-to-noise ratio of 10). |
---|---|
ISSN: | 1061-9348 1608-3199 |
DOI: | 10.1134/S1061934809010079 |