THE IMPACT OF X-RAY UNIT TYPE USED FOR ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURES ON PATIENT DOSES
To investigate whether the X-ray unit type used for interventional endoscopic retrograde cholangiopancreatography (ERCP) procedures may affect patient radiation doses. A total of 471 ERCP procedures performed in 4 hospitals with 4 types of X-ray units were studied. Kerma-area product (KAP), fluorosc...
Saved in:
Published in: | Radiation protection dosimetry Vol. 171; no. 4; pp. 503 - 508 |
---|---|
Main Authors: | , , , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
England
01-12-2016
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To investigate whether the X-ray unit type used for interventional endoscopic retrograde cholangiopancreatography (ERCP) procedures may affect patient radiation doses. A total of 471 ERCP procedures performed in 4 hospitals with 4 types of X-ray units were studied. Kerma-area product (KAP), fluoroscopy time (T) and total number of radiographs acquired (F) were recorded. KAP, T and F values exhibited a great variation, ranging from 0.1 to 130.2 Gy cm
(mean 16 Gy cm
), 0.13 to 33.7 min (mean 5.4 min) and 0 to 26 radiographs (mean 3.5), respectively. The respective mean values for the four types of X-ray units that were investigated were as follows: KAP: 17.4, 12.5, 5.6 and 36.3 Gy cm
, T: 4.7, 5.2, 3.8 and 11.5 min and F: 1.7, 7.4, 1.9 and 4.6 radiographs. The type of the X-ray unit seems to significantly affect patient radiation dose, with the C-arm delivering the lowest and the angiography unit the highest patient doses. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/rpd/ncv465 |