Evaluation of cardiac implantable electronic device lead parameters before and after radiotherapy

Introduction and aim Radiotherapy in cancer patients with a pacemaker or an implantable cardioverter defibrillator might damage the device and possibly result in malfunction of the device. We aimed to examine device malfunction and change in lead parameters in this group. Methods The medical records...

Full description

Saved in:
Bibliographic Details
Published in:Pacing and clinical electrophysiology Vol. 47; no. 9; pp. 1176 - 1184
Main Authors: Larsen, Mia Bang, Jakobsen, Annette Ross, Lundbye‐Christensen, Søren, Riahi, Sam, Thøgersen, Anna Margrethe, Zaremba, Tomas
Format: Journal Article
Language:English
Published: United States Wiley Subscription Services, Inc 01-09-2024
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Introduction and aim Radiotherapy in cancer patients with a pacemaker or an implantable cardioverter defibrillator might damage the device and possibly result in malfunction of the device. We aimed to examine device malfunction and change in lead parameters in this group. Methods The medical records of pacemaker/implantable cardioverter defibrillator patients who underwent radiotherapy at Aalborg University Hospital between July 1, 2016, and June 30, 2021, were reviewed. Treatment was planned according to risk of device malfunction, assessed using local guidelines. Bootstrap linear regression was used to identify possible differences in lead parameters depending on beam energy and anatomical location. Results One hundred nine patients with a pacemaker/implantable cardioverter defibrillator who received a total of 122 radiotherapy courses in the study period were identified. No device malfunctions or significant changes in lead parameters during radiotherapy were observed. Only the left ventricle lead threshold and impedance were found to be significantly different when compared by beam energy; however, these changes were relatively small and unlikely to be of clinical relevance. Conclusion No device malfunction or clinically relevant changes in lead parameters were identified in this study, suggesting that radiotherapy of pacemaker/implantable cardioverter defibrillator patients can be regarded as safe when following relevant safety precautions.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0147-8389
1540-8159
1540-8159
DOI:10.1111/pace.15048