Acute Pancreatitis Due to Clipping of the Ampulla With Over-The-Scope Clip as a Complication of Bleeding Duodenal Ulcer Treatment

Over-the-scope clips (OTSC) (Ovesco, Tübingen, Germany) are commonly used for closure of bowel perforations, fistulas and to achieve hemostasis. This device is attached to the endoscope and delivers a variety of clips, based on diameter and depth, that works through tissue approximation. Complicatio...

Full description

Saved in:
Bibliographic Details
Published in:Curēus (Palo Alto, CA) Vol. 12; no. 7; p. e8963
Main Authors: Sehmbey, Gurbir, Srinivasan, Indu, Chuang, Keng-Yu
Format: Journal Article
Language:English
Published: United States Cureus Inc 02-07-2020
Cureus
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Over-the-scope clips (OTSC) (Ovesco, Tübingen, Germany) are commonly used for closure of bowel perforations, fistulas and to achieve hemostasis. This device is attached to the endoscope and delivers a variety of clips, based on diameter and depth, that works through tissue approximation. Complications including local inflammation, ulcers, or obstruction can occur. When the clip is misplaced or OTSC-associated complications occur, OTSC removal may be indicated. We present a case of a patient who presented to our hospital with upper gastrointestinal (GI) bleeding. OTSC was used to achieve hemostasis, however, the clip was misplaced over the ampulla of Vater. remOVE system (Ovesco, Tübingen, Germany) was used to remove the misplaced clip.