GMI-based biosensor for the detection and quantification of doxorubicin anticancer drugs labeled to Fe3O4 superparamagnetic nanoparticles
We present the results of research on a Co-based giant magnetoimpedance (GMI) biosensor for Fe3O4-labelled doxorubicin (DOX) anticancer drugs. Co62Fe5Ni4Si15B14 ribbons were used as biosensing material for test solutions containing different concentrations of DOX. The plain ribbon, ribbon+Fe3O4 and...
Saved in:
Published in: | Sensors and actuators. A. Physical. Vol. 373; p. 115400 |
---|---|
Main Authors: | , , |
Format: | Journal Article |
Language: | English |
Published: |
Elsevier B.V
01-08-2024
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We present the results of research on a Co-based giant magnetoimpedance (GMI) biosensor for Fe3O4-labelled doxorubicin (DOX) anticancer drugs. Co62Fe5Ni4Si15B14 ribbons were used as biosensing material for test solutions containing different concentrations of DOX. The plain ribbon, ribbon+Fe3O4 and ribbon+DOX2 samples exhibited low coercivity (Hc), high saturation magnetization (Ms) and low anisotropy (Hk). The GMI ratio (ΔZ/Z) increased from 33% to 38% with the increase of doxorubicin content in the ribbon. It was found that the ΔZ/Z ratio first increased rapidly up to a frequency of ∼2 MHz, and then decreased slowly as the frequency increased. While the magnetic field dependent ΔZ/Z has a single peak at 1 MHz, it has a double peak at 5 MHz. It was found that the peak height of ΔZ/Z in the ribbons with test solution of 20 ng/ml and 50 ng/ml increased by ∼3% and ∼6%, respectively, compared to the ribbon without test solution. Then the ΔZ/Z ratio decreased very rapidly with the increase of the applied external magnetic fields. The frequency-dependent GMI sensitivity (η) increased with increasing frequency and reached a maximum value at a critical frequency (∼1.6 MHz), after which it decreased. The detection sensitivity of the GMI biosensor (ξ) increased rapidly with increasing doxorubicin concentration up to ∼80 ng/ml and then remained almost unchanged. The results show that doxorubicin (DOX+Fe3O4), used as an anticancer drug, can be effectively detected even at low concentrations using the GMI based sensors fabricated in this study.
[Display omitted]
•Amorphous Co62Fe5Ni4Si15B14 ribbons were produced by melt spinning for biosensing and the ribbons were heat treated at 340 °C to reduce transverse anisotropy.•The anticancer drug Doxorubicin (DOX) was labeled with Fe3O4 nanoparticles at different concentrations for testing (20 ng/ml and 50 ng/ml).•At 1 MHz, the peak height of the magnetic field dependent impedance curve increased by ∼3% and ∼6% in the samples containing 20 ng/ml and 50 ng/ml test solution, respectively.•The detection sensitivity of the GMI biosensor (ξ) increased rapidly with increasing doxorubicin concentration up to ∼80 ng/ml and then remained almost unchanged•As a result of adding DOX to the ribbons, especially at low concentrations, remarkable increases were obtained in both frequency-dependent impedance values and magnetoimpedance values. |
---|---|
ISSN: | 0924-4247 1873-3069 |