MANAGEMENT OF FUNCTIONAL SPRINT FIDELIS IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR LEADS AT CARDIAC RESYNCHRONIZATION THERAPY-DEFIBRILLATOR GENERATOR REPLACEMENT: AN ALTERNATIVE APPROACH FOR PREVENTION OF INAPPROPRIATE SHOCKS CAUSED BY LEAD FAILURE

Results During a follow-up of 3±2 years after CRT-D generator replacement and switching of Fidelis lead pace/sense pin with bipolar LV lead, in-office interrogations and remote monitoring showed normal device function including biventricular pacing, ventricular tachyarrhythmia detection and appropri...

Full description

Saved in:
Bibliographic Details
Published in:Journal of the American College of Cardiology Vol. 67; no. 13; p. 705
Main Authors: Zhu, Dennis, Chu, Matthew M, Lavell, Kristen A, House, Chad
Format: Journal Article
Language:English
Published: New York Elsevier Inc 05-04-2016
Elsevier Limited
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Results During a follow-up of 3±2 years after CRT-D generator replacement and switching of Fidelis lead pace/sense pin with bipolar LV lead, in-office interrogations and remote monitoring showed normal device function including biventricular pacing, ventricular tachyarrhythmia detection and appropriate antitachycardia therapy.
ISSN:0735-1097
1558-3597
DOI:10.1016/S0735-1097(16)30706-9