Coincidence summing corrections factors calculated for volume 152Eu sources in γ-ray spectromtery using Monte Carlo techniques
True coincidence summing (TCS) correction factors were determined for 152 Eu volume sources used for γ-ray efficiency calibration as the ratio of full energy peak efficiency ( ε ) simulated by MCNP5 code to the apparent full energy peak efficiency ( ε app ) determined experimentally. The results wer...
Saved in:
Published in: | Journal of radioanalytical and nuclear chemistry Vol. 324; no. 2; pp. 687 - 696 |
---|---|
Main Authors: | , , |
Format: | Journal Article |
Language: | English |
Published: |
Cham
Springer International Publishing
2020
Springer Nature B.V |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | True coincidence summing (TCS) correction factors were determined for
152
Eu volume sources used for γ-ray efficiency calibration as the ratio of full energy peak efficiency (
ε
) simulated by MCNP5 code to the apparent full energy peak efficiency (
ε
app
) determined experimentally. The results were compared to those calculated by TrueCoinc software and then validated using certified environmental radioactive samples. The dependance of
152
Eu and
133
Ba TCS factors on volume sample and matrix was investigated. Finally, as our n-type HPGe detector is very sensitive to low energy photons, the contribution of the γ–X coincidence summing to TCS factors for
152
Eu and
133
Ba sources in volume soil samples was studied. |
---|---|
ISSN: | 0236-5731 1588-2780 |
DOI: | 10.1007/s10967-020-07099-4 |