Active decompression improves the haemodynamic state during cardiopulmonary resuscitation
OBJECTIVE--To examine whether use of the active compression-decompression device improves the haemodynamics of cardiopulmonary resuscitation compared with those of conventional cardiopulmonary resuscitation. DESIGN--Prospective crossover study. SETTING--The accident and emergency department of a uni...
Saved in:
Published in: | British Heart Journal Vol. 73; no. 4; pp. 372 - 376 |
---|---|
Main Authors: | , |
Format: | Journal Article |
Language: | English |
Published: |
London
BMJ Publishing Group Ltd and British Cardiovascular Society
01-04-1995
BMJ BMJ Publishing Group LTD |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | OBJECTIVE--To examine whether use of the active compression-decompression device improves the haemodynamics of cardiopulmonary resuscitation compared with those of conventional cardiopulmonary resuscitation. DESIGN--Prospective crossover study. SETTING--The accident and emergency department of a university teaching hospital. PATIENTS--36 adult patients with non-traumatic, normothermic, out of hospital cardiac arrest. INTERVENTIONS--Cardiopulmonary resuscitation was performed during resuscitation in alternating 3 min cycles of conventional and active compression-decompression cardiopulmonary resuscitation. MAIN OUTCOME MEASURES--The end tidal carbon dioxide (ETCO2), femoral arterial pressure, and acid-base analysis of central venous blood measured during the last 30 s of each 3 minute cardiopulmonary resuscitation cycle. RESULTS--ETCO2 was monitored in 36 patients during conventional and active compression-decompression cardiopulmonary resuscitation. Active compression-decompression cardiopulmonary resuscitation caused a significant increase in ETCO2 (P < 0.0002), indicating improved cardiac output. Arterial pressure measurement was carried out in 10 patients. Systolic pressure was significantly greater with active compression-decompression than conventional cardiopulmonary resuscitation (P < 0.007). Central venous blood was taken for acid-base analysis in 11 patients. There was a significant increase in the central venous hydrogen ion concentration (P = 0.025) with rises in the partial pressures of carbon dioxide and oxygen, suggesting improved venous return. CONCLUSIONS--This study confirms that active compression-decompression cardiopulmonary resuscitation is associated with better haemodynamic status than conventional resuscitation. |
---|---|
Bibliography: | local:heartjnl;73/4/372 PMID:7756073 ark:/67375/NVC-GQZ6QQXC-Z href:heartjnl-73-372.pdf istex:7A1DB4F73C461A38C6F78A5D714664279041494B ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |