A parallel plate chamber for calibration of 192 Ir LDR and HDR sources

A good program of quality assurance requires the user to check the calibration of brachytherapy sources. Conventional well-type ionization chambers have restrictions on the length of the source which can be used due to the variation in the relative ionization with distance from the position of maxim...

Full description

Saved in:
Bibliographic Details
Published in:Medical physics (Lancaster) Vol. 26; no. 11; pp. 2438 - 2440
Main Authors: Das, K. R., Kenny, M. B.
Format: Journal Article
Language:English
Published: 01-11-1999
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A good program of quality assurance requires the user to check the calibration of brachytherapy sources. Conventional well-type ionization chambers have restrictions on the length of the source which can be used due to the variation in the relative ionization with distance from the position of maximum sensitivity. We have fabricated a parallel plate chamber, which can be used for both LDR and HDR sources of 192 Ir placed centrally inside. The electrodes are made from printed circuit boards. Sources of lengths up to 15 cm can be calibrated with this chamber. It has a good linear response and stability. It costs little and can be used with any electrometer. The chamber, which has a volume of about 40 cm 3 , has a response of 820 nC  min −1 per cGy m 2  h −1 .