A parallel plate chamber for calibration of 192 Ir LDR and HDR sources
A good program of quality assurance requires the user to check the calibration of brachytherapy sources. Conventional well-type ionization chambers have restrictions on the length of the source which can be used due to the variation in the relative ionization with distance from the position of maxim...
Saved in:
Published in: | Medical physics (Lancaster) Vol. 26; no. 11; pp. 2438 - 2440 |
---|---|
Main Authors: | , |
Format: | Journal Article |
Language: | English |
Published: |
01-11-1999
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A good program of quality assurance requires the user to check the calibration of brachytherapy sources. Conventional well-type ionization chambers have restrictions on the length of the source which can be used due to the variation in the relative ionization with distance from the position of maximum sensitivity. We have fabricated a parallel plate chamber, which can be used for both LDR and HDR sources of
192
Ir
placed centrally inside. The electrodes are made from printed circuit boards. Sources of lengths up to 15 cm can be calibrated with this chamber. It has a good linear response and stability. It costs little and can be used with any electrometer. The chamber, which has a volume of about 40
cm
3
,
has a response of 820
nC
min
−1
per
cGy m
2
h
−1
. |
---|---|
ISSN: | 0094-2405 2473-4209 |
DOI: | 10.1118/1.598762 |