Calibration System Established at TBTAK UME for Infrared Ear Thermometers (IRETs)

The traceable calibration of infrared ear thermometers, which have been used intensively in the health field, has always been issue in Turkey. Moreover, the accuracy of measurements depends entirely on the information determined by the device manufacturer or suppliers on the exception of traceabilit...

Full description

Saved in:
Bibliographic Details
Published in:2018 IEEE International Symposium on Medical Measurements and Applications (MeMeA) pp. 1 - 6
Main Authors: Durmus, Huseyin Okan, Karaboce, Baki, Pehlivan, Ozlem, Korkmaz, Micahit, Kalemci, Murat
Format: Conference Proceeding
Language:English
Published: IEEE 01-06-2018
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Abstract The traceable calibration of infrared ear thermometers, which have been used intensively in the health field, has always been issue in Turkey. Moreover, the accuracy of measurements depends entirely on the information determined by the device manufacturer or suppliers on the exception of traceability chain of our National Metrology Institute (TÜBİTAK UME). However, many years ago, intensive and systematic studies were conducted in America, European Union and Japan to discipline this area. A recent study has also been initiated in Turkey at TÜBİTAK UME which aims to develop a reference calibration system for Infrared (IR) Ear Thermometers. This reference system consists of custom design stirred calibration bath together with blackbody cavities in accordance with both EN and ASTM standards which can be considered as basic standards describing the requirements for the IR ear thermometer calibration system in detail. With the use of the system installed, the calibrations of the infrared ear thermometers will now be able to be performed in our institute.
AbstractList The traceable calibration of infrared ear thermometers, which have been used intensively in the health field, has always been issue in Turkey. Moreover, the accuracy of measurements depends entirely on the information determined by the device manufacturer or suppliers on the exception of traceability chain of our National Metrology Institute (TÜBİTAK UME). However, many years ago, intensive and systematic studies were conducted in America, European Union and Japan to discipline this area. A recent study has also been initiated in Turkey at TÜBİTAK UME which aims to develop a reference calibration system for Infrared (IR) Ear Thermometers. This reference system consists of custom design stirred calibration bath together with blackbody cavities in accordance with both EN and ASTM standards which can be considered as basic standards describing the requirements for the IR ear thermometer calibration system in detail. With the use of the system installed, the calibrations of the infrared ear thermometers will now be able to be performed in our institute.
Author Karaboce, Baki
Kalemci, Murat
Korkmaz, Micahit
Durmus, Huseyin Okan
Pehlivan, Ozlem
Author_xml – sequence: 1
  givenname: Huseyin Okan
  surname: Durmus
  fullname: Durmus, Huseyin Okan
  organization: Medical Metrology Laboratory, TÜBİTAK, National Metrology Institute (UME), Kocaeli, Turkey
– sequence: 2
  givenname: Baki
  surname: Karaboce
  fullname: Karaboce, Baki
  organization: Medical Metrology Laboratory, TÜBİTAK, National Metrology Institute (UME), Kocaeli, Turkey
– sequence: 3
  givenname: Ozlem
  surname: Pehlivan
  fullname: Pehlivan, Ozlem
  organization: Temperature Laboratory, TÜBİTAK, National Metrology Institute (UME), Kocaeli, Turkey
– sequence: 4
  givenname: Micahit
  surname: Korkmaz
  fullname: Korkmaz, Micahit
  organization: Temperature Laboratory, TÜBİTAK, National Metrology Institute (UME), Kocaeli, Turkey
– sequence: 5
  givenname: Murat
  surname: Kalemci
  fullname: Kalemci, Murat
  organization: Temperature Laboratory, TÜBİTAK, National Metrology Institute (UME), Kocaeli, Turkey
BookMark eNotj01LxDAURSMo6IzzB3STpS5akyZtkmUdqhaniFrXw2vzwlT6IUk38--n4Kzu5Vw4cFfkcpxGJOSOs5hzZp4qrDCPE8Z1rKXQmeYXZMXTpQlhkuSabEL4ZYwlmZZKsBvyuYW-azzM3TTS72OYcaBFmKHpu3BAS2Gm9XOdv9OfqqBu8rQcnQe_LAV4Wh_QD9OAM_pAH8qvog6Pt-TKQR9wc841qV-KevsW7T5ey22-izrD5ihrTaOYA8kaoxLdWGmVZMkCU6ekstC2ClOwxrYNipZrdM5osGigtRJTsSb3_9oOEfd_vhvAH_fn0-IE9b9PmQ
ContentType Conference Proceeding
DBID 6IE
6IL
CBEJK
RIE
RIL
DOI 10.1109/MeMeA.2018.8438681
DatabaseName IEEE Electronic Library (IEL) Conference Proceedings
IEEE Proceedings Order Plan All Online (POP All Online) 1998-present by volume
IEEE Xplore All Conference Proceedings
IEEE Electronic Library Online
IEEE Proceedings Order Plans (POP All) 1998-Present
DatabaseTitleList
Database_xml – sequence: 1
  dbid: RIE
  name: IEEE Electronic Library Online
  url: http://ieeexplore.ieee.org/Xplore/DynWel.jsp
  sourceTypes: Publisher
DeliveryMethod fulltext_linktorsrc
EISBN 1538633922
9781538633922
EndPage 6
ExternalDocumentID 8438681
Genre orig-research
GroupedDBID 6IE
6IF
6IL
6IN
AAJGR
ABLEC
ALMA_UNASSIGNED_HOLDINGS
BEFXN
BFFAM
BGNUA
BKEBE
BPEOZ
CBEJK
IEGSK
OCL
RIE
RIL
ID FETCH-LOGICAL-i90t-6c9b70fa40b9728bd4d74029b75f747dacc7e5ad9dcbe3c18eff98ade9acd4e53
IEDL.DBID RIE
IngestDate Thu Jun 29 18:39:48 EDT 2023
IsPeerReviewed false
IsScholarly false
Language English
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-i90t-6c9b70fa40b9728bd4d74029b75f747dacc7e5ad9dcbe3c18eff98ade9acd4e53
PageCount 6
ParticipantIDs ieee_primary_8438681
PublicationCentury 2000
PublicationDate 2018-June
PublicationDateYYYYMMDD 2018-06-01
PublicationDate_xml – month: 06
  year: 2018
  text: 2018-June
PublicationDecade 2010
PublicationTitle 2018 IEEE International Symposium on Medical Measurements and Applications (MeMeA)
PublicationTitleAbbrev MeMeA
PublicationYear 2018
Publisher IEEE
Publisher_xml – name: IEEE
SSID ssj0002684730
Score 1.7092063
Snippet The traceable calibration of infrared ear thermometers, which have been used intensively in the health field, has always been issue in Turkey. Moreover, the...
SourceID ieee
SourceType Publisher
StartPage 1
SubjectTerms ASTM Blackbody cavity
Calibration
Cavity resonators
Infrared Ear Thermometers (IRETs)
Stability criteria
Temperature measurement
Thermal stability
TUBITAK UME
Title Calibration System Established at TBTAK UME for Infrared Ear Thermometers (IRETs)
URI https://ieeexplore.ieee.org/document/8438681
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
link http://sdu.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV1LSwMxEA62J08qrfgmBw8Kbpt2kyY5Vt3SIhUfK3grk2QCHrqVbfv_TXZrRfDiLQRCYCaPmcn3fSHkEoTqp-hlIpCLhAuG4RxElnA1MEJai6mOROHxq3x8V_dZlMm52XJhELECn2EnNqu3fLew61gq6yqeqkHkWTekVjVXa1tPiaolYbV-82KY7k5xisMI3lKdzcBfP6hUF8ho739T75P2DxOPPm3vmAOyg0WLPEdClaldR2vJcZqFKK8GuDsKK5rf5sMH-jbNaAhK6aTwZQSa0wxKGhZGOV_MIwxmSa8mL1m-vG6TfJTld-Nk8zVC8qHZKhlYbSTzwJnRsq-M406GRDB0Ch_yAwfWShTgtLMGU9tT6L1W4FCDdRxFekiaxaLAI0KFd0Jyb8GkKbdeA4AEcKznwsiwYY9JK1pj9lmLX8w2hjj5u_uU7EaD11iqM9JclWs8J42lW19U7voC0xiXiA
link.rule.ids 310,311,782,786,791,792,798,27936,54770
linkProvider IEEE
linkToHtml http://sdu.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwlV1LTwIxEG4UD3pSA8a3PXjQxIVCW9oeUZdAYImPNfFGuu008cBiFvj_truIMfHirWnSNJnpY2b6fV8RutZcdig4EXFgPGKcgD8HgURMdjMujAGqAlF48Com7_IxDjI5dxsuDACU4DNohmb5lm_nZhVKZS3JqOwGnvUOZ0KQiq21qagE3RK_Xr-ZMUS1EkigF-Bbsrke-usPlfIK6e__b_ID1Pjh4uGnzS1ziLYgr6PnQKnKKufhSnQcxz7OqyDuFuslTu_T3gi_JTH2YSke5q4IUHMc6wL7pVHM5rMAhFngm-FLnC5uGyjtx-nDIFp_jhB9KLKMukZlgjjNSKZER2aWWeFTQd_Jnc8QrDZGANdWWZMBNW0JzimpLShtLANOj1Atn-dwjDB3lgvmjM4oZcYprbXQ2pK29SP9lj1B9WCN6WclfzFdG-L07-4rtDtIk_F0PJyMztBeMH6FrDpHtWWxggu0vbCry9J1X_ISmtM
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Abook&rft.genre=proceeding&rft.title=2018+IEEE+International+Symposium+on+Medical+Measurements+and+Applications+%28MeMeA%29&rft.atitle=Calibration+System+Established+at+TBTAK+UME+for+Infrared+Ear+Thermometers+%28IRETs%29&rft.au=Durmus%2C+Huseyin+Okan&rft.au=Karaboce%2C+Baki&rft.au=Pehlivan%2C+Ozlem&rft.au=Korkmaz%2C+Micahit&rft.date=2018-06-01&rft.pub=IEEE&rft.spage=1&rft.epage=6&rft_id=info:doi/10.1109%2FMeMeA.2018.8438681&rft.externalDocID=8438681