System for Non-Contact Diagnostics of Tremor

The paper addresses the problem of creating a digital system for objective control of tremor parameters. A portable tremor recording module has been developed, which utilizes a capacitive proximity sensor based on the capacitance-to-frequency conversion. Experimental studies have shown that the sens...

Full description

Saved in:
Bibliographic Details
Published in:2023 Systems and Technologies of the Digital HealthCare (STDH) pp. 167 - 170
Main Authors: Melnik, Olga V., Nikiforov, Mikhail B., Novikov, Sergey V.
Format: Conference Proceeding
Language:English
Published: IEEE 04-10-2023
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The paper addresses the problem of creating a digital system for objective control of tremor parameters. A portable tremor recording module has been developed, which utilizes a capacitive proximity sensor based on the capacitance-to-frequency conversion. Experimental studies have shown that the sensor's resolution is sufficient to obtain an informative signal for further processing by computational algorithms. The system can be used for both diagnosis of neurological disorders and for the assessment of a person's psychophysical state, particularly for pre-trip monitoring of transportation drivers.