Robust Testing of Displays using the Extreme Linear Dynamic Range CAOS Camera

Proposed and demonstrated for the first time is robust testing of optical displays using the extreme linear Dynamic Range (DR) CAOS camera. Experiments highlight accurate and repeatable CAOS camera-based testing of standard 8-bit (i.e., 48 dB DR) and modified DR 10-bit (i.e., 60 dB DR) computer Liqu...

Full description

Saved in:
Bibliographic Details
Published in:2019 IEEE 2nd British and Irish Conference on Optics and Photonics (BICOP) pp. 1 - 4
Main Authors: Riza, Nabeel A., Ashraf, Nazim, Mazhar, Mohsin A.
Format: Conference Proceeding
Language:English
Published: IEEE 01-12-2019
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Proposed and demonstrated for the first time is robust testing of optical displays using the extreme linear Dynamic Range (DR) CAOS camera. Experiments highlight accurate and repeatable CAOS camera-based testing of standard 8-bit (i.e., 48 dB DR) and modified DR 10-bit (i.e., 60 dB DR) computer Liquid Crystal Displays (LCDs). Results are compared with CMOS camera and light meter-based LCD testing highlighting the robustness of the CAOS camera readings.
DOI:10.1109/BICOP48819.2019.9059583