Is the determination of specific IgE against components using ISAC 112 a reproducible technique?

The ImmunoCAP ISAC 112 is a fluoro-immunoassay that allows detection of specific IgE to 112 molecular components from 51 allergenic sources. We studied the reliability of this technique intra- and inter- assay, as well as inter-batch- and inter-laboratory-assay. Twenty samples were studied, nineteen...

Full description

Saved in:
Bibliographic Details
Published in:PloS one Vol. 9; no. 2; p. e88394
Main Authors: Martínez-Aranguren, Rubén, Lizaso, María T, Goikoetxea, María J, García, Blanca E, Cabrera-Freitag, Paula, Trellez, Oswaldo, Sanz, María L
Format: Journal Article
Language:English
Published: United States Public Library of Science 06-02-2014
Public Library of Science (PLoS)
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The ImmunoCAP ISAC 112 is a fluoro-immunoassay that allows detection of specific IgE to 112 molecular components from 51 allergenic sources. We studied the reliability of this technique intra- and inter- assay, as well as inter-batch- and inter-laboratory-assay. Twenty samples were studied, nineteen sera from polysensitized allergic patients, and the technique calibrator provided by the manufacturer (CTR02). We measured the sIgE from CTR02 and three patients' sera ten times in the same and in different assays. Furthermore, all samples were tested in two laboratories and with two batches of ISAC kit. To evaluate the accuracy of ISAC 112, we contrasted the determinations of CTR02 calibrator with their expected values by T Student test. To analyse the precision, we calculated the coefficient of variation (CV) of the 15 allergens that generate the calibration curve, and to analyse the repeatability and the reproducibility, we calculated the intraclass coefficient correlation (ICC) to each allergen. The results obtained for CTR02 were similar to those expected in 7 of 15 allergens that generate the calibration curve, whereas in 8 allergens the results showed significant differences. The mean CV obtained in the CTR02 determinations was of 9.4%, and the variability of sera from patients was of 22.9%. The agreement in the intra- and inter-assay analysis was very good to 94 allergens and good to one. In the inter-batch analyse, we obtained a very good agreement to 82 allergens, good to 14, moderate to 5 allergens, poor to one, and bad to 1 allergen. In the inter-laboratory analyse, we obtained a very good agreement to 73 allergens, good to 22, moderate to 6 and poor to two allergens. The allergen microarray immunoassay, ISAC 112, is a repeatable and reproducible in vitro diagnostic tool for determination of sIgE beyond the own laboratory.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
Competing Interests: Thermofisher supplies part of the material (ISAC CRD 112 kits) required for this study. This does not alter the authors' adherence to all the PLOS ONE policies on sharing data and materials.
Conceived and designed the experiments: MTL MJG BEG MLS. Performed the experiments: RMMA MTL PCF. Analyzed the data: RMMA MTL MJG OT. Contributed reagents/materials/analysis tools: MTL BEG OT MLS. Wrote the paper: RMMA MJG BEG PCF MLS.
ISSN:1932-6203
1932-6203
DOI:10.1371/journal.pone.0088394