Programmed Atrial Sensitivity: A Critical Determinant in Atrial Fibrillation Detection and Optimal Automatic Mode Switching
Automatic mode switching (AMS) prevents tracking of paroxysmal atria] fibrillation (AF) in dual chamber pacing. The correct detection of AF can be affected by the programmed atrial sensitivity (AS). We prospectively studied the relationship between AS, AF under‐sensing, an d AMS, using unfiltered bi...
Saved in:
Published in: | Pacing and clinical electrophysiology Vol. 21; no. 11; pp. 2214 - 2219 |
---|---|
Main Authors: | , , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
Oxford, UK
Blackwell Publishing Ltd
01-11-1998
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Automatic mode switching (AMS) prevents tracking of paroxysmal atria] fibrillation (AF) in dual chamber pacing. The correct detection of AF can be affected by the programmed atrial sensitivity (AS). We prospectively studied the relationship between AS, AF under‐sensing, an d AMS, using unfiltered bipolar in tracardiac atrial electrogram s recorded from 17 patients during sinus rhythm (SR) and in AF. Overall, 780 rhythms were recorded and replayed onto three dual chamber pacemaker models using different AMS algorithms (Thera DR 7940, Marathon DDDB 294–09, and Meta DDDH 1254), and the ventricular responses were measured. AS was randomly programmed in steps from the highest available AS to half of the mean atrial P wave amplitude (PWA), and the percentage of appropriate AMS responses (defined as a ventricular pacing rate at the expected AMS mode) were recorded. AMS efficacy was related to the programmed AS settings in an exponential manner. At low AS settings, a higher percentage of tests were associated with absence of, or with intermittent AMS and tracking of AF, whereas at higher AS, oversensing of noise during SR occurred. An optimal AS measured approximately 1.3 mV, representing about one‐third of the PWA measured during SR, although oversensing of SR and undersensing of AF continued to occur in 14% of tests and time, respectively, due to the high variation in PWA during AF. Thus, a fixed AS cannot eliminate AF undersensing without inviting noise oversensing, suggesting the need for automatic adjustments of AS, or the use of a rate‐limiting algorithm to prevent rate oscillation during intermittent AF sensing. In conclusion, AMS functions of existing pacemakers were significantly limited by the undersensing of AF and oversensing of noise. Proper adjustment of the AS is important to enable effective AMS during AF. |
---|---|
Bibliography: | ark:/67375/WNG-MJDKCX94-7 ArticleID:PACE2214 istex:100E2A0C0E311760B610E7446BBB84264FFF5D68 ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0147-8389 1540-8159 |
DOI: | 10.1111/j.1540-8159.1998.tb01155.x |