Importance of the choice of the collimator for the detection of small lesions in scintimammography: a phantom study
99mTc methoxyisobutylisonitrile planar scintimammography (SMM) is mostly performed using low-energy high-resolution (LEHR) parallel collimators. We studied whether using a different collimator could improve the detection of small (< 1.5 cm) lesions for which SMM sensitivity is poor. Thirty four b...
Saved in:
Published in: | Physics in medicine & biology Vol. 46; no. 5; p. 1343 |
---|---|
Main Authors: | , , , , |
Format: | Journal Article |
Language: | English |
Published: |
England
01-05-2001
|
Subjects: | |
Online Access: | Get more information |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Abstract | 99mTc methoxyisobutylisonitrile planar scintimammography (SMM) is mostly performed using low-energy high-resolution (LEHR) parallel collimators. We studied whether using a different collimator could improve the detection of small (< 1.5 cm) lesions for which SMM sensitivity is poor. Thirty four breast phantom configurations were considered, either with hot spheres simulating lesions or without any spheres. For each configuration, four planar acquisitions were performed using LEHR, low-energy ultra high-resolution (LEUHR), high-resolution fan-beam (HRFB) and ultra high-resolution fan-beam (UHRFB) collimators. Images corresponding to the 20% and 10% energy windows and to the Jaszczak subtraction were calculated. A database including 156 borderline images was derived. After training, 10 observers scored the images for the presence of a sphere. The performances in sphere detection were studied using receiver operating characteristic (ROC) analysis. For all types of image, the area under the ROC curve was highest with the UHRFB collimator and lowest with either the LEUHR or the HRFB collimator. For the 10% energy window images conventionally used in SMM, the detection sensitivities averaged 91%, 73%, 60% and 55% for the UHRFB, LEHR, HRFB and LEUHR collimators respectively, for the same specificity of 64%. We conclude that detection of small tumours in planar SMM might be significantly improved by using a UHRFB collimator instead of an LEHR collimator. |
---|---|
AbstractList | 99mTc methoxyisobutylisonitrile planar scintimammography (SMM) is mostly performed using low-energy high-resolution (LEHR) parallel collimators. We studied whether using a different collimator could improve the detection of small (< 1.5 cm) lesions for which SMM sensitivity is poor. Thirty four breast phantom configurations were considered, either with hot spheres simulating lesions or without any spheres. For each configuration, four planar acquisitions were performed using LEHR, low-energy ultra high-resolution (LEUHR), high-resolution fan-beam (HRFB) and ultra high-resolution fan-beam (UHRFB) collimators. Images corresponding to the 20% and 10% energy windows and to the Jaszczak subtraction were calculated. A database including 156 borderline images was derived. After training, 10 observers scored the images for the presence of a sphere. The performances in sphere detection were studied using receiver operating characteristic (ROC) analysis. For all types of image, the area under the ROC curve was highest with the UHRFB collimator and lowest with either the LEUHR or the HRFB collimator. For the 10% energy window images conventionally used in SMM, the detection sensitivities averaged 91%, 73%, 60% and 55% for the UHRFB, LEHR, HRFB and LEUHR collimators respectively, for the same specificity of 64%. We conclude that detection of small tumours in planar SMM might be significantly improved by using a UHRFB collimator instead of an LEHR collimator. |
Author | Le Cloirec, J Buvat, I Di Paola, R Laffont, S Bourguet, P |
Author_xml | – sequence: 1 givenname: I surname: Buvat fullname: Buvat, I email: buvat@imed.jussieu.fr organization: U494 INSERM, CHU Pitié-Salpêtrière, Paris, France. buvat@imed.jussieu.fr – sequence: 2 givenname: S surname: Laffont fullname: Laffont, S – sequence: 3 givenname: J surname: Le Cloirec fullname: Le Cloirec, J – sequence: 4 givenname: P surname: Bourguet fullname: Bourguet, P – sequence: 5 givenname: R surname: Di Paola fullname: Di Paola, R |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/11384056$$D View this record in MEDLINE/PubMed |
BookMark | eNpNj8lOw0AQROcQRBb4AQ5ofsB49oy5oYglUiQucI5mMzaaxfJMDv57HLGIQ6tUpdetrjVYxBQdADcY3WEkZY0QxVWDOa-ZqHlNEV6A1V-4BOucPxHCWBJ2CZYYU8kQFyuQ92FIY1HROJhaWDoHTZf6fy553wdV0gjbec6RdcWZ0qd4ZnJQ3kPv8uwz7CPMpo9l3gghfYxq6KZ7qODQqVhSgLmc7HQFLlrls7v-0Q14f3p8271Uh9fn_e7hUBkmeKmEoWhriSAMS8tpKykjW6yFcUxTTYUmtDGUIKK0arg0nGjp1IwYg2hrG7IBt993h5MOzh6HcX5rnI6_7ckXv-hdsg |
CitedBy_id | crossref_primary_10_1007_s12149_014_0820_2 crossref_primary_10_1007_s41365_016_0088_4 crossref_primary_10_1097_MNM_0000000000000053 crossref_primary_10_1088_0031_9155_61_14_5390 crossref_primary_10_1155_2016_7462014 crossref_primary_10_1016_j_nima_2006_08_034 crossref_primary_10_1088_1748_0221_7_01_P01011 crossref_primary_10_1109_TNS_2015_2484070 crossref_primary_10_1016_j_pbiomolbio_2012_03_003 crossref_primary_10_1097_MNM_0000000000001400 crossref_primary_10_1109_TNS_2006_879824 |
ContentType | Journal Article |
DBID | CGR CUY CVF ECM EIF NPM |
DOI | 10.1088/0031-9155/46/5/301 |
DatabaseName | Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed |
DatabaseTitle | MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) |
DatabaseTitleList | MEDLINE |
Database_xml | – sequence: 1 dbid: ECM name: MEDLINE url: https://search.ebscohost.com/login.aspx?direct=true&db=cmedm&site=ehost-live sourceTypes: Index Database |
DeliveryMethod | no_fulltext_linktorsrc |
Discipline | Medicine Biology Physics |
ExternalDocumentID | 11384056 |
Genre | Journal Article |
GroupedDBID | --- -DZ -~X .GJ .HR 02O 123 1JI 1PV 1WK 29O 3O- 4.4 53G 5B3 5RE 5VS 5ZH 5ZI 7.M 7.Q 9BW AAGCD AAGCF AAJIO AAJKP AALHV AATNI ABCXL ABHWH ABJNI ABLJU ABQJV ABTAH ABVAM ACAFW ACGFS ACHIP ACMRT ACWPO AEFHF AENEX AERVB AFFNX AFYNE AHSEE AKPSB ALMA_UNASSIGNED_HOLDINGS AOAED ASPBG ATQHT AVWKF AZFZN BBWZM CBCFC CEBXE CGR CJUJL CRLBU CS3 CUY CVF DU5 EBS ECM EDWGO EIF EJD EMSAF EPQRW EQZZN F5P FEDTE HAK HVGLF H~9 IHE IJHAN IOP IZVLO J5H JCGBZ KOT LAP M45 N5L N9A NPM NT- NT. P2P PJBAE Q02 R4D RIN RKQ RNS RO9 ROL RPA S3P SY9 T37 TN5 UCJ W28 X7L XPP ZGI ZMT ZXP ZY4 |
ID | FETCH-LOGICAL-c465t-6c307d262418d53f834271b6ce4b3b36b239c3202aba958c52b8ea271cc03fd92 |
ISSN | 0031-9155 |
IngestDate | Wed Oct 16 00:50:27 EDT 2024 |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 5 |
Language | English |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-c465t-6c307d262418d53f834271b6ce4b3b36b239c3202aba958c52b8ea271cc03fd92 |
PMID | 11384056 |
ParticipantIDs | pubmed_primary_11384056 |
PublicationCentury | 2000 |
PublicationDate | 2001-05-01 |
PublicationDateYYYYMMDD | 2001-05-01 |
PublicationDate_xml | – month: 05 year: 2001 text: 2001-05-01 day: 01 |
PublicationDecade | 2000 |
PublicationPlace | England |
PublicationPlace_xml | – name: England |
PublicationTitle | Physics in medicine & biology |
PublicationTitleAlternate | Phys Med Biol |
PublicationYear | 2001 |
SSID | ssj0011824 |
Score | 1.7513802 |
Snippet | 99mTc methoxyisobutylisonitrile planar scintimammography (SMM) is mostly performed using low-energy high-resolution (LEHR) parallel collimators. We studied... |
SourceID | pubmed |
SourceType | Index Database |
StartPage | 1343 |
SubjectTerms | Breast Neoplasms - diagnostic imaging Female Humans Image Processing, Computer-Assisted Mammography Phantoms, Imaging Radionuclide Imaging - instrumentation Radiopharmaceuticals ROC Curve Sensitivity and Specificity Technetium Tc 99m Sestamibi |
Title | Importance of the choice of the collimator for the detection of small lesions in scintimammography: a phantom study |
URI | https://www.ncbi.nlm.nih.gov/pubmed/11384056 |
Volume | 46 |
hasFullText | |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://sdu.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV3dT9swELda0CZe0PjYgG3ID7xNEU3suAlv0BWBBNMkmMQb8ldYJZJWaovEf8-d7aQpiAke9hIl58RJfb-e78O-I-RAsL7qG64imF9sxKXMo0xLFWmhVGqE0tbgRuGzq_6vm-znkA87nbrI2oL2XzkNNOA17px9B7ebToEA58BzOALX4fgmvp-XTqPG_2sI_4OAG7WuMAt3iaZ2s8LQ2JnVteY4LTFYfW-n9RpzmCOrGTwBP8Fnt_b7oyd_sfxw2cpPG1Rct6ZUu0frwL3DV0j31Nj_8wc5W3LbXsgCt4gv-WMv7I_B_RjF8lII6wRG4m7u4yi_lxwX8WKZYC2MWRxhevq2MA7-yFE73O0ka8x8OqcXIh_EpFsjGTrDWk7ojsAT5l_X4vqkdGyPYwaGrc9o_u_WZ8m466Yu6YJqhdr34LIJWoGp5pN-h08Je7TgAw8b2iEXWG295yoUhc6eWTROs7n-RNaDSUKPPZY2SMdWm-SDL1L6uEk-XgYuAjHwdotMFzCj44ICiqiHWXPVwIwCzBypgRne42BGA8zoqKIvYHZEJQ0gow5k2-TP6fB6cBaF-h2R5iKdRULDBGISAUpiZlJWZIwn_VgJbbliigmVsFyzpJdIJfM002miMivhFq17rDB58pmsVOPK7hBqi4IzjjQwVwwo0aB1cmtlz5hcGJPuki9-CG8nPknLbT24e6-2fCVrC1h-I6sFSAD7nXSnZr7vGPsEc-p43w |
link.rule.ids | 782 |
linkProvider | EBSCOhost |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Importance+of+the+choice+of+the+collimator+for+the+detection+of+small+lesions+in+scintimammography%3A+a+phantom+study&rft.jtitle=Physics+in+medicine+%26+biology&rft.au=Buvat%2C+I&rft.au=Laffont%2C+S&rft.au=Le+Cloirec%2C+J&rft.au=Bourguet%2C+P&rft.date=2001-05-01&rft.issn=0031-9155&rft.volume=46&rft.issue=5&rft.spage=1343&rft_id=info:doi/10.1088%2F0031-9155%2F46%2F5%2F301&rft_id=info%3Apmid%2F11384056&rft_id=info%3Apmid%2F11384056&rft.externalDocID=11384056 |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0031-9155&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0031-9155&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0031-9155&client=summon |