Accuracy of alveolar bone measurements from cone beam computed tomography acquired using varying settings

Structured Objectives To investigate the accuracy and reliability of cone beam computed tomography (CBCT) measurements of buccal alveolar bone height (BBH) and thickness (BBT) using custom acquisition settings. Settings and Sample Population School of Dentistry, Oregon Health & Science Universit...

Full description

Saved in:
Bibliographic Details
Published in:Orthodontics & craniofacial research Vol. 18; no. S1; pp. 127 - 136
Main Authors: Cook, V. C., Timock, A. M., Crowe, J. J., Wang, M., Covell Jr, D. A.
Format: Journal Article
Language:English
Published: England Blackwell Publishing Ltd 01-04-2015
Wiley Subscription Services, Inc
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Structured Objectives To investigate the accuracy and reliability of cone beam computed tomography (CBCT) measurements of buccal alveolar bone height (BBH) and thickness (BBT) using custom acquisition settings. Settings and Sample Population School of Dentistry, Oregon Health & Science University. Twelve embalmed cadavers. Materials and Methods Cadaver heads were imaged by CBCT (i‐CAT® 17–19, Imaging Sciences International, Hatfield, PA) using a ‘long scan’ (LS) setting with 619 projection images, 360° revolution, 26.9 s duration, and 0.2 mm voxel size, and using a ‘short scan’ (SS) setting with 169 projection images, 180° rotation, 4.8 s duration, and 0.3 mm voxel size. BBH and BBT were measured with 65 teeth, indirectly from CBCT images and directly through dissection. Comparisons were assessed using paired t‐tests (p ≤ 0.05). Level of agreement was assessed by concordance correlation coefficients, Pearson's correlation coefficients, and Bland–Altman plots. Results Mean differences in measurements compared to direct measurements were as follows, LS 0.17 ± 0.12 (BBH) and 0.10 ± 0.07 mm (BBT), and SS 0.41 ± 0.32 (BBH) and 0.12 ± 0.11 mm (BBT). No statistical differences were found with any of BBH or BBT measurements. Correlation coefficients and Bland–Altman plots showed agreement was high between direct and indirect measurement methods, although agreement was stronger for measurements of BBH than BBT. Conclusions Compared to the LS, the similarity in results with the reduced scan times and hence reduced effective radiation dose, favors use of shorter scans, unless other purposes for higher resolution imaging can be defined.
Bibliography:ArticleID:OCR12072
istex:2BAEABE4034E6587B1B1E58F24BBA0CF5EB67E11
ark:/67375/WNG-ZHJFBX9Z-N
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1601-6335
1601-6343
DOI:10.1111/ocr.12072