Conditioning data for calculation of the modulation transfer function
A method for conditioning data used in the measurement of the modulation transfer function (MTF) is discussed. This method is based upon imposing the constraint that the edge spread function (ESF) is monotonic. The advantages of this technique, when applicable, are demonstrated with simulated exampl...
Saved in:
Published in: | Medical physics (Lancaster) Vol. 30; no. 2; pp. 248 - 253 |
---|---|
Main Authors: | , |
Format: | Journal Article |
Language: | English |
Published: |
United States
American Association of Physicists in Medicine
01-02-2003
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A method for conditioning data used in the measurement of the modulation transfer function (MTF) is discussed. This method is based upon imposing the constraint that the edge spread function (ESF) is monotonic. The advantages of this technique, when applicable, are demonstrated with simulated examples for which the true MTF is known. The application of this technique in the measurement of the MTF of a digital detector in clinical use is also demonstrated. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0094-2405 2473-4209 |
DOI: | 10.1118/1.1534111 |