Impact of implementing a category 1 external quality assurance scheme for monitoring harmonization of clinical laboratories in Spain

The objective of the present study was to examine the evolution of the analytical performance specifications (APS) used in External Quality Assurance (EQA) schemes, as well as the efficacy of a category 1 EQA scheme in monitoring the harmonization of clinical laboratory results in Spain. A review of...

Full description

Saved in:
Bibliographic Details
Published in:Advances in laboratory medicine Vol. 1; no. 1; p. 20200008
Main Authors: Ricós, Carmen, Fernández-Calle, Pilar, Marqués, Fernando, Minchinela, Joana, Salas, Ángel, Martínez-Bru, Cecília, Boned, Beatriz, Gómez-Rioja, Rubén, Cortés, Marià, González-Lao, Elisabet, García-Lario, J V, Tejedor-Ganduxé, Xavier, Bullich, Sandra, Ventura, Montse, Simón, Margarida, Vilaplana, Carlos, González-Tarancón, Ricardo, Fernández-Fernández, Mª Pilar, Ramón-Bauzá, Francisco, Corte, Zoraida, Llopis, Mª Antonia, Díaz-Garzón, Jorge, Perich, Carmen
Format: Journal Article
Language:English
Published: Germany De Gruyter 01-03-2020
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The objective of the present study was to examine the evolution of the analytical performance specifications (APS) used in External Quality Assurance (EQA) schemes, as well as the efficacy of a category 1 EQA scheme in monitoring the harmonization of clinical laboratory results in Spain. A review of the literature on the types of quality specifications used in schemes in other countries and their evolution was performed. In addition, a comparative analysis of the potential impact that different APS from eight countries had on clinical decision-making was made based on three measurands: sodium, thyroid-stimulating hormone (TSH), and activated partial thromboplastin time (aPTT). Harmonization of analytical methods was demonstrated by assessing whether average results deviated from the certified reference value of control materials within the APS derived from biological variation (BV). The APS used in EQA have evolved from state-of-the-art models to BV. Poor clinical decision-making would occur if the results accepted by some APS were applied. In Spain, only 2 of the 18 measurands studied are considered to be well harmonized. Closer collaboration between laboratories and analytical system providers would be required to resolve discrepancies.