Berlin Heart EXCOR Ventricular Assist Device: Multilayer Membrane Rupture in a Pediatric Patient

A 2-year-old child was implanted with an Berlin Heart EXCOR Ventricular Assist Device (Berlin Heart, Berlin, Germany) as a bridge to heart transplantation for idiopathic dilated cardiomyopathy. At postoperative day 296, a significant reduction of membrane movement was observed. The device was explan...

Full description

Saved in:
Bibliographic Details
Published in:The Annals of thoracic surgery Vol. 102; no. 2; pp. e129 - e130
Main Authors: Di Molfetta, Arianna, PhD, Filippelli, Sergio, PhD, Ferrari, Gianfranco, PhD, Secinaro, Aurelio, PhD, Zielinski, Krystzof, PhD, Amodeo, Antonio, MD
Format: Journal Article
Language:English
Published: Netherlands Elsevier Inc 01-08-2016
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A 2-year-old child was implanted with an Berlin Heart EXCOR Ventricular Assist Device (Berlin Heart, Berlin, Germany) as a bridge to heart transplantation for idiopathic dilated cardiomyopathy. At postoperative day 296, a significant reduction of membrane movement was observed. The device was explanted and tested on a hydronumerical circulation simulator. Findings suggested that the integrity of the multilayered membrane had been compromised. This was confirmed by a computed tomography scan of the device. The computed tomography evidenced a detachment of the 3-layered membrane, with a thinner, convex layer on the side of the air chamber and an opposite convexity of the remaining membranes. These showed an additional air space within the layers.
Bibliography:ObjectType-Case Study-2
SourceType-Scholarly Journals-1
ObjectType-Feature-4
content type line 23
ObjectType-Report-1
ObjectType-Article-3
ISSN:0003-4975
1552-6259
DOI:10.1016/j.athoracsur.2016.01.022