Berlin Heart EXCOR Ventricular Assist Device: Multilayer Membrane Rupture in a Pediatric Patient
A 2-year-old child was implanted with an Berlin Heart EXCOR Ventricular Assist Device (Berlin Heart, Berlin, Germany) as a bridge to heart transplantation for idiopathic dilated cardiomyopathy. At postoperative day 296, a significant reduction of membrane movement was observed. The device was explan...
Saved in:
Published in: | The Annals of thoracic surgery Vol. 102; no. 2; pp. e129 - e130 |
---|---|
Main Authors: | , , , , , |
Format: | Journal Article |
Language: | English |
Published: |
Netherlands
Elsevier Inc
01-08-2016
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A 2-year-old child was implanted with an Berlin Heart EXCOR Ventricular Assist Device (Berlin Heart, Berlin, Germany) as a bridge to heart transplantation for idiopathic dilated cardiomyopathy. At postoperative day 296, a significant reduction of membrane movement was observed. The device was explanted and tested on a hydronumerical circulation simulator. Findings suggested that the integrity of the multilayered membrane had been compromised. This was confirmed by a computed tomography scan of the device. The computed tomography evidenced a detachment of the 3-layered membrane, with a thinner, convex layer on the side of the air chamber and an opposite convexity of the remaining membranes. These showed an additional air space within the layers. |
---|---|
Bibliography: | ObjectType-Case Study-2 SourceType-Scholarly Journals-1 ObjectType-Feature-4 content type line 23 ObjectType-Report-1 ObjectType-Article-3 |
ISSN: | 0003-4975 1552-6259 |