Interference of Electronic Apex Locators with Implantable Cardioverter Defibrillators
Abstract Introduction The purpose of this in vitro study was to evaluate the potential electromagnetic interference of electronic apex locators (EALs) on implantable cardioverter defibrillators (ICDs). Methods Four different EALs were tested for their ability to interfere with the correct function o...
Saved in:
Published in: | Journal of endodontics Vol. 40; no. 2; pp. 277 - 280 |
---|---|
Main Authors: | , , , |
Format: | Journal Article |
Language: | English |
Published: |
United States
Elsevier Inc
01-02-2014
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Abstract | Abstract Introduction The purpose of this in vitro study was to evaluate the potential electromagnetic interference of electronic apex locators (EALs) on implantable cardioverter defibrillators (ICDs). Methods Four different EALs were tested for their ability to interfere with the correct function of 3 different ICDs. Each ICD was placed in a plastic container with 1.5 L physiological saline, and the EAL unit was placed at a distance of 2.5 cm from the ICD. The file electrode and lip clip were placed directly against the ICD. The EAL was turned on for 30 seconds while continuously showing the “APEX” mark. As a negative control, the ICD was tested without EAL for 30 seconds. An electrosurgical unit served as a positive control. During each test, the ICD output was monitored continuously by real-time telemetry, and after completion of the experiment, intracardiac electrocardiograms were printed. The tests were repeated 3 times for each device. The electrocardiograms were examined for interference on ICD ventricular activity. Results All EALs tested and the negative control failed to produce electromagnetic interference in each of the ICDs tested. The electrosurgical unit induced interference in the ICDs, which were detected as episodes of ventricular tachycardia and led to the initiation of electrical shocks in all ICDs. Conclusions The 4 EALs tested did not interfere with the correct functioning of ICDs in vitro. |
---|---|
AbstractList | The purpose of this in vitro study was to evaluate the potential electromagnetic interference of electronic apex locators (EALs) on implantable cardioverter defibrillators (ICDs).
Four different EALs were tested for their ability to interfere with the correct function of 3 different ICDs. Each ICD was placed in a plastic container with 1.5 L physiological saline, and the EAL unit was placed at a distance of 2.5 cm from the ICD. The file electrode and lip clip were placed directly against the ICD. The EAL was turned on for 30 seconds while continuously showing the “APEX” mark. As a negative control, the ICD was tested without EAL for 30 seconds. An electrosurgical unit served as a positive control. During each test, the ICD output was monitored continuously by real-time telemetry, and after completion of the experiment, intracardiac electrocardiograms were printed. The tests were repeated 3 times for each device. The electrocardiograms were examined for interference on ICD ventricular activity.
All EALs tested and the negative control failed to produce electromagnetic interference in each of the ICDs tested. The electrosurgical unit induced interference in the ICDs, which were detected as episodes of ventricular tachycardia and led to the initiation of electrical shocks in all ICDs.
The 4 EALs tested did not interfere with the correct functioning of ICDs in vitro. INTRODUCTIONThe purpose of this in vitro study was to evaluate the potential electromagnetic interference of electronic apex locators (EALs) on implantable cardioverter defibrillators (ICDs).METHODSFour different EALs were tested for their ability to interfere with the correct function of 3 different ICDs. Each ICD was placed in a plastic container with 1.5 L physiological saline, and the EAL unit was placed at a distance of 2.5 cm from the ICD. The file electrode and lip clip were placed directly against the ICD. The EAL was turned on for 30 seconds while continuously showing the "APEX" mark. As a negative control, the ICD was tested without EAL for 30 seconds. An electrosurgical unit served as a positive control. During each test, the ICD output was monitored continuously by real-time telemetry, and after completion of the experiment, intracardiac electrocardiograms were printed. The tests were repeated 3 times for each device. The electrocardiograms were examined for interference on ICD ventricular activity.RESULTSAll EALs tested and the negative control failed to produce electromagnetic interference in each of the ICDs tested. The electrosurgical unit induced interference in the ICDs, which were detected as episodes of ventricular tachycardia and led to the initiation of electrical shocks in all ICDs.CONCLUSIONSThe 4 EALs tested did not interfere with the correct functioning of ICDs in vitro. Abstract Introduction The purpose of this in vitro study was to evaluate the potential electromagnetic interference of electronic apex locators (EALs) on implantable cardioverter defibrillators (ICDs). Methods Four different EALs were tested for their ability to interfere with the correct function of 3 different ICDs. Each ICD was placed in a plastic container with 1.5 L physiological saline, and the EAL unit was placed at a distance of 2.5 cm from the ICD. The file electrode and lip clip were placed directly against the ICD. The EAL was turned on for 30 seconds while continuously showing the “APEX” mark. As a negative control, the ICD was tested without EAL for 30 seconds. An electrosurgical unit served as a positive control. During each test, the ICD output was monitored continuously by real-time telemetry, and after completion of the experiment, intracardiac electrocardiograms were printed. The tests were repeated 3 times for each device. The electrocardiograms were examined for interference on ICD ventricular activity. Results All EALs tested and the negative control failed to produce electromagnetic interference in each of the ICDs tested. The electrosurgical unit induced interference in the ICDs, which were detected as episodes of ventricular tachycardia and led to the initiation of electrical shocks in all ICDs. Conclusions The 4 EALs tested did not interfere with the correct functioning of ICDs in vitro. The purpose of this in vitro study was to evaluate the potential electromagnetic interference of electronic apex locators (EALs) on implantable cardioverter defibrillators (ICDs). Four different EALs were tested for their ability to interfere with the correct function of 3 different ICDs. Each ICD was placed in a plastic container with 1.5 L physiological saline, and the EAL unit was placed at a distance of 2.5 cm from the ICD. The file electrode and lip clip were placed directly against the ICD. The EAL was turned on for 30 seconds while continuously showing the "APEX" mark. As a negative control, the ICD was tested without EAL for 30 seconds. An electrosurgical unit served as a positive control. During each test, the ICD output was monitored continuously by real-time telemetry, and after completion of the experiment, intracardiac electrocardiograms were printed. The tests were repeated 3 times for each device. The electrocardiograms were examined for interference on ICD ventricular activity. All EALs tested and the negative control failed to produce electromagnetic interference in each of the ICDs tested. The electrosurgical unit induced interference in the ICDs, which were detected as episodes of ventricular tachycardia and led to the initiation of electrical shocks in all ICDs. The 4 EALs tested did not interfere with the correct functioning of ICDs in vitro. |
Author | Brand, Henk S., PhD de Cock, Carel C., MD, PhD Shemesh, Hagay, DDS, PhD Idzahi, Karim, DDS |
Author_xml | – sequence: 1 fullname: Idzahi, Karim, DDS – sequence: 2 fullname: de Cock, Carel C., MD, PhD – sequence: 3 fullname: Shemesh, Hagay, DDS, PhD – sequence: 4 fullname: Brand, Henk S., PhD |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/24461418$$D View this record in MEDLINE/PubMed |
BookMark | eNp9kU1v1DAQhi3Uim4Lf4ADypFLwvgjcSIhpGop7Uor9QA9W449EQ5Ze7GzLf33OGzh0ENPc3neVzPPnJMTHzwS8o5CRYE2H8dqDOgrBpRXICtg8hVZ0Va2Ja9rcUJWAF1XMt51Z-Q8pRGASs7la3LGhGiooO2K3G38jHHAiN5gEYbiakIzx-CdKS73-LvYBqPnEFPx4OYfxWa3n7SfdT9hsdbRunCPMRcUX3BwfXTT9Bd-Q04HPSV8-zQvyN3Xq-_rm3J7e71ZX25LIyidy0Eww8DaehAADbOdbKDlrNc9Qou9yFuyoTMSagttz3qkXNtWDzWg0Lxr-QX5cOzdx_DrgGlWO5cM5i08hkNSVHRMAm9kk1F2RE0MKUUc1D66nY6PioJadKpRLTrVolOBVFlnDr1_6j_0O7T_I__8ZeDTEcB85b3DqJJxi0rrYvaobHAv939-FjeTy-r19BMfMY3hEH32p6hKTIH6tjx0-Sfl2ZdoKf8D8bWdgg |
CitedBy_id | crossref_primary_10_1016_j_joen_2014_12_011 crossref_primary_10_1111_adj_12281 crossref_primary_10_1111_iej_12487 crossref_primary_10_1111_eos_12183 crossref_primary_10_2174_18742106_v16_e221227_2022_67 crossref_primary_10_1016_j_jdent_2015_12_008 crossref_primary_10_5005_jp_journals_10047_0079 crossref_primary_10_1080_00016357_2017_1364419 crossref_primary_10_1055_a_2002_0295 crossref_primary_10_1111_eos_12313 crossref_primary_10_1111_pace_14051 crossref_primary_10_1016_j_joen_2015_02_007 crossref_primary_10_1080_19932820_2018_1547071 crossref_primary_10_1111_aej_12688 crossref_primary_10_3390_app14041445 |
Cites_doi | 10.14219/jada.archive.1975.0565 10.1056/NEJM199912163412503 10.14219/jada.archive.1974.0559 10.1177/00220345620410020801 10.14219/jada.archive.2010.0320 10.14219/jada.archive.2010.0324 10.1002/clc.21997 10.1097/00004770-200212000-00009 10.1111/iej.12000 10.1016/j.joen.2006.03.005 10.1093/eurheartj/sum062 10.1046/j.1460-9592.2002.01496.x 10.1016/j.tripleo.2005.02.016 10.14219/jada.archive.1918.0368 10.1016/S1079-2104(98)90394-8 10.1016/0300-5712(82)90027-6 10.1007/s11606-007-0392-0 10.1016/j.jacc.2012.07.017 10.1016/S0099-2399(96)80097-5 10.14219/jada.archive.2010.0224 10.1038/bdj.2007.1043 |
ContentType | Journal Article |
Copyright | American Association of Endodontists 2014 American Association of Endodontists Copyright © 2014 American Association of Endodontists. Published by Elsevier Inc. All rights reserved. |
Copyright_xml | – notice: American Association of Endodontists – notice: 2014 American Association of Endodontists – notice: Copyright © 2014 American Association of Endodontists. Published by Elsevier Inc. All rights reserved. |
DBID | CGR CUY CVF ECM EIF NPM AAYXX CITATION 7X8 |
DOI | 10.1016/j.joen.2013.07.027 |
DatabaseName | Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed CrossRef MEDLINE - Academic |
DatabaseTitle | MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) CrossRef MEDLINE - Academic |
DatabaseTitleList | MEDLINE - Academic MEDLINE |
Database_xml | – sequence: 1 dbid: ECM name: MEDLINE url: https://search.ebscohost.com/login.aspx?direct=true&db=cmedm&site=ehost-live sourceTypes: Index Database |
DeliveryMethod | fulltext_linktorsrc |
Discipline | Dentistry |
EISSN | 1878-3554 |
EndPage | 280 |
ExternalDocumentID | 10_1016_j_joen_2013_07_027 24461418 S0099239913006481 1_s2_0_S0099239913006481 |
Genre | Journal Article |
GroupedDBID | --- --K .1- .FO 08G 0R~ 1B1 1P~ 1RT 1~5 354 4.4 457 4G. 4Q1 4Q2 4Q3 53G 5GY 5RE 5VS 7-5 8F7 AAEDT AAEDW AAGKA AAKAS AALRI AAQFI AAQXK AAXUO AAYEP ABFRF ABJNI ABLJU ABMAC ABOCM ACGFO ACGFS ADBBV ADBIZ ADMUD ADPAM ADZCM AEFWE AENEX AEVXI AFCTW AFFNX AFJKZ AFRHN AFTJW AFTRI AFUWQ AHHHB AHRYX AITUG AIZYK AJUYK AKRWK ALMA_UNASSIGNED_HOLDINGS AMRAJ ASPBG AVWKF AZFZN BELOY C45 C5W CS3 DU5 EBS EFJIC EJD F5P FDB FEDTE FGOYB G-2 GBLVA HDX HMK HMO HVGLF HZ~ IHE JF9 JG8 KMI KOM LH1 M28 M41 MO0 NQ- NTWIH O9- OAG OAH OB- OL1 OM. OVD OWU OWV OWW OWX OWY OWZ P2P R2- RIG ROL RPZ SAE SEL SES SSZ T8P TEORI UNMZH VVN WUQ XH2 XYM YFH Z5R ZFV ZGI ~S- AAIAV AGZHU ALXNB AWKKM ZA5 CGR CUY CVF ECM EIF NPM AAYXX CITATION 7X8 |
ID | FETCH-LOGICAL-c411t-f42c20dd5f40062d9760832babe08eb44462f9c705d08b2be13ad8af50e4a3983 |
ISSN | 0099-2399 |
IngestDate | Fri Oct 25 06:08:23 EDT 2024 Thu Sep 26 17:02:50 EDT 2024 Sat Sep 28 07:59:40 EDT 2024 Fri Feb 23 02:15:19 EST 2024 Tue Oct 15 22:55:50 EDT 2024 |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 2 |
Keywords | Electromagnetic interference implantable cardioverter defibrillator electronic apex locator |
Language | English |
License | Copyright © 2014 American Association of Endodontists. Published by Elsevier Inc. All rights reserved. |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-c411t-f42c20dd5f40062d9760832babe08eb44462f9c705d08b2be13ad8af50e4a3983 |
Notes | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
PMID | 24461418 |
PQID | 1492703676 |
PQPubID | 23479 |
PageCount | 4 |
ParticipantIDs | proquest_miscellaneous_1492703676 crossref_primary_10_1016_j_joen_2013_07_027 pubmed_primary_24461418 elsevier_sciencedirect_doi_10_1016_j_joen_2013_07_027 elsevier_clinicalkeyesjournals_1_s2_0_S0099239913006481 |
PublicationCentury | 2000 |
PublicationDate | 2014-02-01 |
PublicationDateYYYYMMDD | 2014-02-01 |
PublicationDate_xml | – month: 02 year: 2014 text: 2014-02-01 day: 01 |
PublicationDecade | 2010 |
PublicationPlace | United States |
PublicationPlace_xml | – name: United States |
PublicationTitle | Journal of endodontics |
PublicationTitleAlternate | J Endod |
PublicationYear | 2014 |
Publisher | Elsevier Inc |
Publisher_xml | – name: Elsevier Inc |
References | Beach, Bramwell, Hutter (bib13) 1996; 22 Miller, Leonelli, Latham (bib5) 1998; 85 Wilson, Broberg, Baumgartner (bib16) 2006; 32 Pinski, Trohman (bib2) 2002; 25 Buxton, Lee, Fisher (bib19) 1999; 341 Woolley, Woodworth, Dobbs (bib4) 1974; 89 Custer (bib11) 1918; 5 Sunada (bib12) 1962; 41 Gomez, Duran-Sindreu, Jara Clemente (bib15) 2013; 46 Ector, Vardas (bib22) 2007; 9 Luker (bib10) 1982; 10 Garofalo, Ede, Dorn (bib14) 2002; 28 Patel, Glick, Lessard (bib7) 2005; 99 Simon, Linde, Bennette (bib9) 1975; 91 Crossley, Poole (bib24) 2010; 141 Morita (bib17) 2011 Greenspon, Patel, Lau (bib20) 2012; 60 Brand, Entjes, Nieuw Amerongen (bib8) 2007; 203 Carlson (bib23) 2010; 141 (bib1) 2009 Zhan, Baine, Sedrakyan (bib21) 2008; 23 Roedig, Shah, Elayi (bib6) 2010; 141 VDW Gold [operation manual]. Munich, Germany: VDW; 2010. Misiri, Kusumoto, Goldschlager (bib3) 2012; 35 Roedig (10.1016/j.joen.2013.07.027_bib6) 2010; 141 Ector (10.1016/j.joen.2013.07.027_bib22) 2007; 9 Zhan (10.1016/j.joen.2013.07.027_bib21) 2008; 23 Custer (10.1016/j.joen.2013.07.027_bib11) 1918; 5 Gomez (10.1016/j.joen.2013.07.027_bib15) 2013; 46 Woolley (10.1016/j.joen.2013.07.027_bib4) 1974; 89 Simon (10.1016/j.joen.2013.07.027_bib9) 1975; 91 Miller (10.1016/j.joen.2013.07.027_bib5) 1998; 85 Buxton (10.1016/j.joen.2013.07.027_bib19) 1999; 341 Crossley (10.1016/j.joen.2013.07.027_bib24) 2010; 141 (10.1016/j.joen.2013.07.027_bib1) 2009 Pinski (10.1016/j.joen.2013.07.027_bib2) 2002; 25 Beach (10.1016/j.joen.2013.07.027_bib13) 1996; 22 Brand (10.1016/j.joen.2013.07.027_bib8) 2007; 203 Misiri (10.1016/j.joen.2013.07.027_bib3) 2012; 35 Morita (10.1016/j.joen.2013.07.027_bib17) 2011 Greenspon (10.1016/j.joen.2013.07.027_bib20) 2012; 60 Carlson (10.1016/j.joen.2013.07.027_bib23) 2010; 141 Garofalo (10.1016/j.joen.2013.07.027_bib14) 2002; 28 Sunada (10.1016/j.joen.2013.07.027_bib12) 1962; 41 Patel (10.1016/j.joen.2013.07.027_bib7) 2005; 99 10.1016/j.joen.2013.07.027_bib18 Luker (10.1016/j.joen.2013.07.027_bib10) 1982; 10 Wilson (10.1016/j.joen.2013.07.027_bib16) 2006; 32 |
References_xml | – volume: 85 start-page: 33 year: 1998 end-page: 36 ident: bib5 article-title: Selective interference with pacemaker activity by electrical dental devices publication-title: Oral Surg Oral Med Oral Pathol contributor: fullname: Latham – volume: 141 start-page: 1052 year: 2010 end-page: 1053 ident: bib23 article-title: Pacemakers and dental devices publication-title: J Am Dent Assoc contributor: fullname: Carlson – volume: 35 start-page: 321 year: 2012 end-page: 328 ident: bib3 article-title: Electromagnetic interference and implanted cardiac devices: the medical environment (part II) publication-title: Clin Cardiol contributor: fullname: Goldschlager – volume: 41 start-page: 375 year: 1962 end-page: 387 ident: bib12 article-title: New method for measuring the length of the root canal publication-title: J Dent Res contributor: fullname: Sunada – volume: 32 start-page: 847 year: 2006 end-page: 852 ident: bib16 article-title: Safety of electronic apex locators and pulp testers in patients with implanted cardiac pacemakers or cardioverter/defibrillators publication-title: J Endod contributor: fullname: Baumgartner – volume: 10 start-page: 326 year: 1982 end-page: 332 ident: bib10 article-title: The pacemaker patient in the dental surgery publication-title: J Dent contributor: fullname: Luker – year: 2009 ident: bib1 publication-title: Dental equipment and implantable pacemakers and defibrillators – volume: 22 start-page: 182 year: 1996 end-page: 184 ident: bib13 article-title: Use of an electronic apex locator on a cardiac pacemaker patient publication-title: J Endod contributor: fullname: Hutter – volume: 28 start-page: 831 year: 2002 end-page: 833 ident: bib14 article-title: The effect of electronic apex locators on cardiac pacemaker function publication-title: J Endod contributor: fullname: Dorn – year: 2011 ident: bib17 article-title: Root ZX mini [operation instructions] contributor: fullname: Morita – volume: 5 start-page: 815 year: 1918 end-page: 819 ident: bib11 article-title: Exact methods for locating the apical foramen publication-title: J Natl Dent Assoc contributor: fullname: Custer – volume: 341 start-page: 1882 year: 1999 end-page: 1890 ident: bib19 article-title: A randomized study of the prevention of sudden death in patients with coronary artery disease publication-title: N Engl J Med contributor: fullname: Fisher – volume: 23 start-page: 13 year: 2008 end-page: 19 ident: bib21 article-title: Cardiac device implantation in the United States from 1997 through 2004: a population-based analysis publication-title: J Gen Intern Med contributor: fullname: Sedrakyan – volume: 25 start-page: 1496 year: 2002 end-page: 1509 ident: bib2 article-title: Interference in implanted cardiac devices, part II publication-title: Pacing Clin Electrophysiol contributor: fullname: Trohman – volume: 99 start-page: 430 year: 2005 ident: bib7 article-title: Absence of in vivo effects of dental instruments on pacemaker function publication-title: Oral Surg Oral Med Oral Pathol contributor: fullname: Lessard – volume: 203 start-page: 577 year: 2007 end-page: 579 ident: bib8 article-title: Interference of electrical dental equipment with implantable cardioverter-defibrillators publication-title: Br Dent J contributor: fullname: Nieuw Amerongen – volume: 60 start-page: 1540 year: 2012 end-page: 1545 ident: bib20 article-title: Trends in permanent pacemaker implantation in the United States from 1993 to 2009 publication-title: J Am Coll Cardiol contributor: fullname: Lau – volume: 141 start-page: 521 year: 2010 end-page: 526 ident: bib6 article-title: Interference of cardiac pacemaker and implantable cardioverter-defibrillator activity during electronic dental device use publication-title: J Am Dent Assoc contributor: fullname: Elayi – volume: 141 start-page: 1053 year: 2010 ident: bib24 article-title: More about pacemakers publication-title: J Am Dent Assoc contributor: fullname: Poole – volume: 46 start-page: 399 year: 2013 end-page: 405 ident: bib15 article-title: The effects of six electronic apex locators on pacemaker function: an publication-title: Int Endod J contributor: fullname: Jara Clemente – volume: 9 start-page: I44 year: 2007 end-page: I49 ident: bib22 article-title: Current use of pacemakers, implantable cardioverter defibrillators, and resynchronization devices: data from the registry of the European Heart Rhythm Association publication-title: Eur Heart J Suppl contributor: fullname: Vardas – volume: 89 start-page: 1099 year: 1974 end-page: 1101 ident: bib4 article-title: A preliminary evaluation of the effects of electrical pulp testers on dogs with artificial pacemakers publication-title: J Am Dent Assoc contributor: fullname: Dobbs – volume: 91 start-page: 1224 year: 1975 end-page: 1229 ident: bib9 article-title: The individual with a pacemaker in the dental environment publication-title: J Am Dent Assoc contributor: fullname: Bennette – volume: 91 start-page: 1224 year: 1975 ident: 10.1016/j.joen.2013.07.027_bib9 article-title: The individual with a pacemaker in the dental environment publication-title: J Am Dent Assoc doi: 10.14219/jada.archive.1975.0565 contributor: fullname: Simon – year: 2011 ident: 10.1016/j.joen.2013.07.027_bib17 contributor: fullname: Morita – volume: 341 start-page: 1882 year: 1999 ident: 10.1016/j.joen.2013.07.027_bib19 article-title: A randomized study of the prevention of sudden death in patients with coronary artery disease publication-title: N Engl J Med doi: 10.1056/NEJM199912163412503 contributor: fullname: Buxton – volume: 89 start-page: 1099 year: 1974 ident: 10.1016/j.joen.2013.07.027_bib4 article-title: A preliminary evaluation of the effects of electrical pulp testers on dogs with artificial pacemakers publication-title: J Am Dent Assoc doi: 10.14219/jada.archive.1974.0559 contributor: fullname: Woolley – volume: 41 start-page: 375 year: 1962 ident: 10.1016/j.joen.2013.07.027_bib12 article-title: New method for measuring the length of the root canal publication-title: J Dent Res doi: 10.1177/00220345620410020801 contributor: fullname: Sunada – volume: 141 start-page: 1052 year: 2010 ident: 10.1016/j.joen.2013.07.027_bib23 article-title: Pacemakers and dental devices publication-title: J Am Dent Assoc doi: 10.14219/jada.archive.2010.0320 contributor: fullname: Carlson – volume: 141 start-page: 1053 year: 2010 ident: 10.1016/j.joen.2013.07.027_bib24 article-title: More about pacemakers publication-title: J Am Dent Assoc doi: 10.14219/jada.archive.2010.0324 contributor: fullname: Crossley – volume: 35 start-page: 321 year: 2012 ident: 10.1016/j.joen.2013.07.027_bib3 article-title: Electromagnetic interference and implanted cardiac devices: the medical environment (part II) publication-title: Clin Cardiol doi: 10.1002/clc.21997 contributor: fullname: Misiri – volume: 28 start-page: 831 year: 2002 ident: 10.1016/j.joen.2013.07.027_bib14 article-title: The effect of electronic apex locators on cardiac pacemaker function publication-title: J Endod doi: 10.1097/00004770-200212000-00009 contributor: fullname: Garofalo – volume: 46 start-page: 399 year: 2013 ident: 10.1016/j.joen.2013.07.027_bib15 article-title: The effects of six electronic apex locators on pacemaker function: an in vitro study publication-title: Int Endod J doi: 10.1111/iej.12000 contributor: fullname: Gomez – volume: 32 start-page: 847 year: 2006 ident: 10.1016/j.joen.2013.07.027_bib16 article-title: Safety of electronic apex locators and pulp testers in patients with implanted cardiac pacemakers or cardioverter/defibrillators publication-title: J Endod doi: 10.1016/j.joen.2006.03.005 contributor: fullname: Wilson – ident: 10.1016/j.joen.2013.07.027_bib18 – volume: 9 start-page: I44 year: 2007 ident: 10.1016/j.joen.2013.07.027_bib22 article-title: Current use of pacemakers, implantable cardioverter defibrillators, and resynchronization devices: data from the registry of the European Heart Rhythm Association publication-title: Eur Heart J Suppl doi: 10.1093/eurheartj/sum062 contributor: fullname: Ector – volume: 25 start-page: 1496 year: 2002 ident: 10.1016/j.joen.2013.07.027_bib2 article-title: Interference in implanted cardiac devices, part II publication-title: Pacing Clin Electrophysiol doi: 10.1046/j.1460-9592.2002.01496.x contributor: fullname: Pinski – volume: 99 start-page: 430 year: 2005 ident: 10.1016/j.joen.2013.07.027_bib7 article-title: Absence of in vivo effects of dental instruments on pacemaker function publication-title: Oral Surg Oral Med Oral Pathol doi: 10.1016/j.tripleo.2005.02.016 contributor: fullname: Patel – volume: 5 start-page: 815 year: 1918 ident: 10.1016/j.joen.2013.07.027_bib11 article-title: Exact methods for locating the apical foramen publication-title: J Natl Dent Assoc doi: 10.14219/jada.archive.1918.0368 contributor: fullname: Custer – volume: 85 start-page: 33 year: 1998 ident: 10.1016/j.joen.2013.07.027_bib5 article-title: Selective interference with pacemaker activity by electrical dental devices publication-title: Oral Surg Oral Med Oral Pathol doi: 10.1016/S1079-2104(98)90394-8 contributor: fullname: Miller – volume: 10 start-page: 326 year: 1982 ident: 10.1016/j.joen.2013.07.027_bib10 article-title: The pacemaker patient in the dental surgery publication-title: J Dent doi: 10.1016/0300-5712(82)90027-6 contributor: fullname: Luker – volume: 23 start-page: 13 year: 2008 ident: 10.1016/j.joen.2013.07.027_bib21 article-title: Cardiac device implantation in the United States from 1997 through 2004: a population-based analysis publication-title: J Gen Intern Med doi: 10.1007/s11606-007-0392-0 contributor: fullname: Zhan – volume: 60 start-page: 1540 year: 2012 ident: 10.1016/j.joen.2013.07.027_bib20 article-title: Trends in permanent pacemaker implantation in the United States from 1993 to 2009 publication-title: J Am Coll Cardiol doi: 10.1016/j.jacc.2012.07.017 contributor: fullname: Greenspon – volume: 22 start-page: 182 year: 1996 ident: 10.1016/j.joen.2013.07.027_bib13 article-title: Use of an electronic apex locator on a cardiac pacemaker patient publication-title: J Endod doi: 10.1016/S0099-2399(96)80097-5 contributor: fullname: Beach – year: 2009 ident: 10.1016/j.joen.2013.07.027_bib1 – volume: 141 start-page: 521 year: 2010 ident: 10.1016/j.joen.2013.07.027_bib6 article-title: Interference of cardiac pacemaker and implantable cardioverter-defibrillator activity during electronic dental device use publication-title: J Am Dent Assoc doi: 10.14219/jada.archive.2010.0224 contributor: fullname: Roedig – volume: 203 start-page: 577 year: 2007 ident: 10.1016/j.joen.2013.07.027_bib8 article-title: Interference of electrical dental equipment with implantable cardioverter-defibrillators publication-title: Br Dent J doi: 10.1038/bdj.2007.1043 contributor: fullname: Brand |
SSID | ssj0017337 |
Score | 2.2268188 |
Snippet | Abstract Introduction The purpose of this in vitro study was to evaluate the potential electromagnetic interference of electronic apex locators (EALs) on... The purpose of this in vitro study was to evaluate the potential electromagnetic interference of electronic apex locators (EALs) on implantable cardioverter... The purpose of this in vitro study was to evaluate the potential electromagnetic interference of electronic apex locators (EALs) on implantable cardioverter... INTRODUCTIONThe purpose of this in vitro study was to evaluate the potential electromagnetic interference of electronic apex locators (EALs) on implantable... |
SourceID | proquest crossref pubmed elsevier |
SourceType | Aggregation Database Index Database Publisher |
StartPage | 277 |
SubjectTerms | Defibrillators, Implantable Dentistry Electric Impedance Electrical Equipment and Supplies Electrocardiography Electromagnetic Fields Electromagnetic interference electronic apex locator Electrosurgery - instrumentation Endocrinology & Metabolism Equipment Safety Humans implantable cardioverter defibrillator Materials Testing Odontometry - instrumentation Root Canal Preparation - instrumentation Tooth Apex - anatomy & histology |
Title | Interference of Electronic Apex Locators with Implantable Cardioverter Defibrillators |
URI | https://www.clinicalkey.es/playcontent/1-s2.0-S0099239913006481 https://dx.doi.org/10.1016/j.joen.2013.07.027 https://www.ncbi.nlm.nih.gov/pubmed/24461418 https://search.proquest.com/docview/1492703676 |
Volume | 40 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://sdu.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Lj9MwELa6ywEuiDflpSBxq1LFjvM6Lk1REY8D3ZUQF8tJbJaym642rQT8emZiOwldFQESl6hK3MSZ-TIzHs-DkBdSgiJIZOoXWcl90NfUl6XWfhaxWMkgBbGMPt3FMnn_Mc3nfD4auQrB_bn_ymk4B7zGzNm_4HZ3UzgBv4HncASuw_GP-N66-FzxWAwb7PvcHF2ob5O3oLzaDjutBxaLAwNt2_ypmQlNVRjkCXJIYzIAgAQH7zFhVV3BqrbeDCLmX1c_5OkXm2ZmOjXneWdwVtgkzwhgTHw6m8xa4fQuN76dPvr4VJ2rpvX4LORn-d3eZmfUy0sblblQ9dfJcjq8bB0ZlLvY5144Z5mPqbZD4WxqOVkQsqGktd1fjNJmph3UFX1gXBOr6WqtsNgtDdtKrSzptZ_b8Z_P3lC_YdPAX-JEcB640xdz01Fmpyg3FQ0Tgbgy9IBcYyDxUOB-ij5021lJ6Iq3mje02Vsm0HB3bvsspH0roNYSOr5Fblr-e0cGe7fJSNV3yPUcw86wc-BdcjLEoLfWXo9BDzHoOQx6iEFvgEFviEHvVwzeIyev5sezhW_bd_glp3Tja85KFlRVpDlm6lZg-IK9zwpZKJABBec8ZjorkyCqgrRghaKhrFKpo0BxGWZpeJ8c1utaPSSeijllKkt1nGqOi2xWFJIWsBRnPKx0NiYTRzJxYaq0CBe-uBJIYIEEFkEigMBjkjiqCpd_DBpTNfYzbsQ-9o5J1P3TWqjG8hQAuN8-8bljngDxjXtyslbrLTyJZwxr4CXxmDwwXO3eACxvMJ5p-uif5_uY3Oi_tCfkcHO5VU_JQVNtn7UA_Qk-OsCm |
link.rule.ids | 315,782,786,27933,27934 |
linkProvider | Elsevier |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Interference+of+Electronic+Apex+Locators+with+Implantable+Cardioverter+Defibrillators&rft.jtitle=Journal+of+endodontics&rft.au=Idzahi%2C+Karim%2C+DDS&rft.au=de+Cock%2C+Carel+C.%2C+MD%2C+PhD&rft.au=Shemesh%2C+Hagay%2C+DDS%2C+PhD&rft.au=Brand%2C+Henk+S.%2C+PhD&rft.date=2014-02-01&rft.issn=0099-2399&rft.volume=40&rft.issue=2&rft.spage=277&rft.epage=280&rft_id=info:doi/10.1016%2Fj.joen.2013.07.027&rft.externalDBID=ECK1-s2.0-S0099239913006481&rft.externalDocID=1_s2_0_S0099239913006481 |
thumbnail_m | http://sdu.summon.serialssolutions.com/2.0.0/image/custom?url=https%3A%2F%2Fcdn.clinicalkey.com%2Fck-thumbnails%2F00992399%2FS0099239913X00147%2Fcov150h.gif |