Performance of Photon-Counting Breast Computed Tomography, Digital Mammography, and Digital Breast Tomosynthesis in Evaluating Breast Specimens

Rationale and Objectives This study compared a novel photon-counting breast computed tomography (pcBCT) system with digital mammography (DM) and digital breast tomosynthesis (DBT) systems. For this reason, surgical specimens were examined with all three techniques and rated by three observers. Mater...

Full description

Saved in:
Bibliographic Details
Published in:Academic radiology Vol. 24; no. 2; pp. 184 - 190
Main Authors: Rößler, Ann-Christin, Kalender, Willi, Kolditz, Daniel, Steiding, Christian, Ruth, Veikko, Preuss, Caroline, Peter, Sandra Christina, Brehm, Barbara, Hammon, Matthias, Schulz-Wendtland, Rüdiger, Wenkel, Evelyn
Format: Journal Article
Language:English
Published: United States Elsevier Inc 01-02-2017
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Rationale and Objectives This study compared a novel photon-counting breast computed tomography (pcBCT) system with digital mammography (DM) and digital breast tomosynthesis (DBT) systems. For this reason, surgical specimens were examined with all three techniques and rated by three observers. Materials and Methods A total of 30 surgical specimens were investigated with DM, DBT, and pcBCT; the associated images were shown to three experienced radiologists. Findings (22 microcalcifications and 23 mass lesions) were recorded and compared to the results of the pathological examination. Sensitivity and specificity for detection of microcalcifications and lesions were calculated and displayed using receiver operating characteristic curves. Results Sensitivity for microcalcifications was 82% for DM, 70% for DBT, and 85% for pcBCT. Specificity for microcalcifications was 71% for DM, 75% for DBT, and 83% for pcBCT. Sensitivity for lesions was 45% for DM, 62% for DBT, and 65% for pcBCT. Specificity for lesions was 76% for DM, 62% for DBT, and 76% for pcBCT. Conclusions pcBCT showed a comparable or superior performance compared to the clinically approved DM and DBT systems. Mass lesion detectability can be increased further by the use of contrast media.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1076-6332
1878-4046
DOI:10.1016/j.acra.2016.09.017