Use of region II of the a/q stability diagram for fast scanning of a linear quadrupole mass spectrometer

For evaluation as a detector for high-speed gas chromatography, a linear quadrupole mass spectrometer was operated in rf only mode by using Region II ( a = 0; 7.514 < q < 7.580) of the Mathieu a/q stability diagram. The available power supply and the diameter of the quadrupole rods of the mass...

Full description

Saved in:
Bibliographic Details
Published in:Journal of the American Society for Mass Spectrometry Vol. 8; no. 5; pp. 539 - 544
Main Authors: Grimm, Casey C., Clawson, Richard, Short, Robert T.
Format: Journal Article
Language:English
Published: New York, NY Elsevier Inc 01-05-1997
Elsevier Science
Springer Nature B.V
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:For evaluation as a detector for high-speed gas chromatography, a linear quadrupole mass spectrometer was operated in rf only mode by using Region II ( a = 0; 7.514 < q < 7.580) of the Mathieu a/q stability diagram. The available power supply and the diameter of the quadrupole rods of the mass spectrometer placed an upper mass limit of ~ m/z 93. Scan rates of 1000 scans/s were obtained with mass spectral peaks resolved over an 80-u range The m/z 91 and 92 ions produced from the electron ionization of toluene are resolved with an R 1/2 of 135. A potential difference between the source and the quadruple mass filter of up to 1000 V was used to accelerate ions into the quadrupole. Broadening of mass-to-charge ratio peaks results from the time constant of the signal amplification rather than the small number of rf cycles the ions experience. The expected loss of sensitivity relative to Region I is observed, and the problem of mass aliasing is discussed.
ISSN:1044-0305
1879-1123
DOI:10.1016/S1044-0305(97)00006-8