Fracture of the central polyethylene tibial spine in posterior stabilized total knee arthroplasty

Between 1993 and 1998, 332 I-BII (Insall Burstein II) prostheses were implanted in our department. We studied 5 cases of tibial polyethylene spine fractures. The fractures occurred 3 to 5 years following the surgery. All cases with fractures had mild medial laxity that was noted on follow-up of 2 to...

Full description

Saved in:
Bibliographic Details
Published in:The Journal of arthroplasty Vol. 18; no. 5; pp. 672 - 674
Main Authors: Hendel, David, Garti, Avraham, Weisbort, Moshe
Format: Journal Article
Language:English
Published: United States Elsevier Inc 01-08-2003
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Between 1993 and 1998, 332 I-BII (Insall Burstein II) prostheses were implanted in our department. We studied 5 cases of tibial polyethylene spine fractures. The fractures occurred 3 to 5 years following the surgery. All cases with fractures had mild medial laxity that was noted on follow-up of 2 to 3 years following arthroplasty. One also developed mild hyperextension. All the patients had a revision surgery. In 2 cases, the polyethylene was changed to a thicker one, together with medial reinforcement. Three cases were revised to a rotating hinge prosthesis.