Fracture of the central polyethylene tibial spine in posterior stabilized total knee arthroplasty
Between 1993 and 1998, 332 I-BII (Insall Burstein II) prostheses were implanted in our department. We studied 5 cases of tibial polyethylene spine fractures. The fractures occurred 3 to 5 years following the surgery. All cases with fractures had mild medial laxity that was noted on follow-up of 2 to...
Saved in:
Published in: | The Journal of arthroplasty Vol. 18; no. 5; pp. 672 - 674 |
---|---|
Main Authors: | , , |
Format: | Journal Article |
Language: | English |
Published: |
United States
Elsevier Inc
01-08-2003
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Between 1993 and 1998, 332 I-BII (Insall Burstein II) prostheses were implanted in our department. We studied 5 cases of tibial polyethylene spine fractures. The fractures occurred 3 to 5 years following the surgery. All cases with fractures had mild medial laxity that was noted on follow-up of 2 to 3 years following arthroplasty. One also developed mild hyperextension. All the patients had a revision surgery. In 2 cases, the polyethylene was changed to a thicker one, together with medial reinforcement. Three cases were revised to a rotating hinge prosthesis. |
---|---|
Bibliography: | ObjectType-Case Study-2 SourceType-Scholarly Journals-1 ObjectType-Feature-4 content type line 23 ObjectType-Report-1 ObjectType-Article-3 |
ISSN: | 0883-5403 1532-8406 |
DOI: | 10.1016/S0883-5403(03)00192-X |