False low pulse oximetry reading associated with the concomitant use of a peripheral nerve stimulator and an evoked-potential stimulator
One of the sources of error in pulse oximetry readings is associated with an abnormal signal-to-noise ratio. The pulse oximeter distinguishes the light absorbance of arterial blood from that of other absorbers by differentiating between a constant component and a pulsating component. The pulsating c...
Saved in:
Published in: | Journal of clinical anesthesia Vol. 9; no. 7; pp. 591 - 596 |
---|---|
Main Authors: | , , |
Format: | Journal Article |
Language: | English |
Published: |
United States
Elsevier Inc
01-11-1997
|
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Abstract | One of the sources of error in pulse oximetry readings is associated with an abnormal signal-to-noise ratio. The pulse oximeter distinguishes the light absorbance of arterial blood from that of other absorbers by differentiating between a constant component and a pulsating component. The pulsating component is almost exclusively the result of arteriolar bed pulsations. Because pulse oximetry is based on the assumption that arterial blood is the only pulsatile absorber, any other fluctuating phenomenon could constitute a source of error. We report a case in which a low pulse oximetry reading was associated with concomitant use of a pulse oximeter and a peripheral nerve stimulator on the same arm. Further tests conducted using a nerve stimulator and a sensory evoked potential stimulator with different amplitudes and frequencies confirmed the association and delineated the relationship between frequency and amplitude of stimulation and the degree of artificial desaturation. A theoretical explanation for this phenomenon is presented. |
---|---|
AbstractList | One of the sources of error in pulse oximetry readings is associated with an abnormal signal-to-noise ratio. The pulse oximeter distinguishes the light absorbance of arterial blood from that of other absorbers by differentiating between a constant component and a pulsating component. The pulsating component is almost exclusively the result of arteriolar bed pulsations. Because pulse oximetry is based on the assumption that arterial blood is the only pulsatile absorber, any other fluctuating phenomenon could constitute a source of error. We report a case in which a low pulse oximetry reading was associated with concomitant use of a pulse oximeter and a peripheral nerve stimulator on the same arm. Further tests conducted using a nerve stimulator and a sensory evoked potential stimulator with different amplitudes and frequencies confirmed the association and delineated the relationship between frequency and amplitude of stimulation and the degree of artificial desaturation. A theoretical explanation for this phenomenon is presented. |
Author | Gravenstein, Dietrich Keidan, Ilan Sidi, Avner |
Author_xml | – sequence: 1 givenname: Ilan surname: Keidan fullname: Keidan, Ilan – sequence: 2 givenname: Avner surname: Sidi fullname: Sidi, Avner – sequence: 3 givenname: Dietrich surname: Gravenstein fullname: Gravenstein, Dietrich |
BackLink | https://www.ncbi.nlm.nih.gov/pubmed/9347439$$D View this record in MEDLINE/PubMed |
BookMark | eNqFkc1u3CAUhVGVKpmkfYRIrKJ24RbsMeauoirKT6VIXbRdIwzXHVobHMCT5A362LFnRumyCwTofOdeuOeUHPngkZBzzj5xxsXn7wzqspBcsg_QfGSMzzd4Q1ZcNlWxrks4IqtX5IScpvSbMTYL_JgcQ7Vu1hWsyN8b3SekfXik47ScwpMbMMdnGlFb539RnVIwTme09NHlDc0bpCZ4EwaXtc90Wkwd1XTE6MYNRt1Tj3GLNGU3TL3OIVLt7bwobsMftMUYMvrsZvAf8o687ZanvD_sZ-TnzfWPq7vi_tvt16sv94WpBMtFiUa3ci24AF3L2patbYFZaBE4lFBz2ZXC8FJWVnRWAkjWlbo2poHG6EpWZ-RiX3eM4WHClNXgksG-1x7DlFQDlRBiB9Z70MSQUsROjdENOj4rztSSgNoloJbxKmjULgEFs-_80GBqB7SvrsPIZ_1yr-P8y63DqJJx6A1aF9FkZYP7T4cXO6GazQ |
CitedBy_id | crossref_primary_10_1111_j_1365_2044_2004_03921_x crossref_primary_10_1097_00001503_199811000_00009 |
Cites_doi | 10.1097/00000542-198901000-00019 10.1097/00000542-199610000-00012 10.1007/BF01617328 10.1007/BF01616745 10.1213/00000539-199608000-00012 |
ContentType | Journal Article |
Copyright | 1997 |
Copyright_xml | – notice: 1997 |
DBID | CGR CUY CVF ECM EIF NPM AAYXX CITATION 7X8 |
DOI | 10.1016/S0952-8180(97)00152-9 |
DatabaseName | Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed CrossRef MEDLINE - Academic |
DatabaseTitle | MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) CrossRef MEDLINE - Academic |
DatabaseTitleList | MEDLINE - Academic MEDLINE |
Database_xml | – sequence: 1 dbid: ECM name: MEDLINE url: https://search.ebscohost.com/login.aspx?direct=true&db=cmedm&site=ehost-live sourceTypes: Index Database |
DeliveryMethod | fulltext_linktorsrc |
EISSN | 1873-4529 |
EndPage | 596 |
ExternalDocumentID | 10_1016_S0952_8180_97_00152_9 9347439 S0952818097001529 |
Genre | Journal Article Case Reports |
GroupedDBID | --- --K --M -RU .1- .FO .GJ .~1 0R~ 1B1 1P~ 1RT 1~. 1~5 29K 3O- 3V. 4.4 457 4G. 53G 5GY 5RE 5VS 7-5 71M 7RV 7X7 88E 8FI 8FJ 8G5 8P~ 9JM AACTN AAEDT AAEDW AAIAV AAIKJ AAKOC AALRI AAOAW AAQFI AAQXK AAWTL AAXUO ABBQC ABFRF ABJNI ABLVK ABMAC ABMZM ABUWG ABXDB ABYKQ ACDAQ ACGFO ACGFS ACIUM ACRLP ADBBV ADEZE ADFRT ADMUD AEBSH AEFWE AEKER AENEX AEVXI AFCTW AFKRA AFKWA AFRHN AFTJW AFXIZ AGHFR AGUBO AGYEJ AHHHB AHMBA AIEXJ AIKHN AITUG AJBFU AJOXV AJRQY AJUYK ALMA_UNASSIGNED_HOLDINGS AMFUW AMRAJ ANZVX ASPBG AVWKF AXJTR AZFZN AZQEC BENPR BKEYQ BKOJK BLXMC BNPGV BPHCQ BVXVI CCPQU CS3 D-I DU5 DWQXO EBS EFJIC EFLBG EJD EMOBN EO8 EO9 EP2 EP3 EX3 F5P FDB FEDTE FGOYB FIRID FNPLU FYGXN FYUFA G-2 G-Q GBLVA GNUQQ GUQSH HDV HMCUK HMK HMO HVGLF HZ~ IHE J1W KOM LCYCR LX1 M1P M29 M2O M41 MO0 N4W N9A NAPCQ O-L O9- O90 OAUVE OL- OZT P-8 P-9 P2P PC. PQQKQ PROAC PSQYO Q38 R2- RIG ROL RPZ SAE SCC SDF SDG SDP SEL SES SEW SPCBC SSH SSZ SV3 T5K UKHRP UNMZH WOW WUQ XPP Z5R ~G- AAXKI AFJKZ AKRWK ALIPV CGR CUY CVF ECM EIF NPM AAYXX CITATION 7X8 |
ID | FETCH-LOGICAL-c360t-2ecab846169a585d2bdb90d9be91929518f26c1283d6fd89980f2a5cc797ca383 |
ISSN | 0952-8180 |
IngestDate | Fri Oct 25 21:11:59 EDT 2024 Thu Nov 21 22:38:04 EST 2024 Sat Sep 28 07:39:47 EDT 2024 Fri Feb 23 02:17:01 EST 2024 |
IsPeerReviewed | true |
IsScholarly | true |
Issue | 7 |
Keywords | Equipment evoked potential stimulators nerve stimulators pulse monitors |
Language | English |
LinkModel | OpenURL |
MergedId | FETCHMERGED-LOGICAL-c360t-2ecab846169a585d2bdb90d9be91929518f26c1283d6fd89980f2a5cc797ca383 |
Notes | ObjectType-Case Study-2 SourceType-Scholarly Journals-1 ObjectType-Feature-4 content type line 23 ObjectType-Report-1 ObjectType-Article-3 |
PMID | 9347439 |
PQID | 79366638 |
PQPubID | 23479 |
PageCount | 6 |
ParticipantIDs | proquest_miscellaneous_79366638 crossref_primary_10_1016_S0952_8180_97_00152_9 pubmed_primary_9347439 elsevier_sciencedirect_doi_10_1016_S0952_8180_97_00152_9 |
PublicationCentury | 1900 |
PublicationDate | 1997-11-01 |
PublicationDateYYYYMMDD | 1997-11-01 |
PublicationDate_xml | – month: 11 year: 1997 text: 1997-11-01 day: 01 |
PublicationDecade | 1990 |
PublicationPlace | United States |
PublicationPlace_xml | – name: United States |
PublicationTitle | Journal of clinical anesthesia |
PublicationTitleAlternate | J Clin Anesth |
PublicationYear | 1997 |
Publisher | Elsevier Inc |
Publisher_xml | – name: Elsevier Inc |
References | Wukitsch, Tobler, Pologe, Peterson (BIB4) 1988; 4 Block, Stahl (BIB3) 1995; 11 Dumas, Wahr, Tremper (BIB5) 1996; 83 Pologe (BIB2) 1987; vol 25 Tremper, Barker (BIB1) 1989; 70 Barker, Shah (BIB6) 1996; 85 Pologe (10.1016/S0952-8180(97)00152-9_BIB2) 1987; vol 25 Block (10.1016/S0952-8180(97)00152-9_BIB3) 1995; 11 Barker (10.1016/S0952-8180(97)00152-9_BIB6) 1996; 85 Dumas (10.1016/S0952-8180(97)00152-9_BIB5) 1996; 83 Tremper (10.1016/S0952-8180(97)00152-9_BIB1) 1989; 70 Wukitsch (10.1016/S0952-8180(97)00152-9_BIB4) 1988; 4 |
References_xml | – volume: 70 start-page: 98 year: 1989 end-page: 108 ident: BIB1 article-title: Pulse oximetry publication-title: Anesthesiology contributor: fullname: Barker – volume: 85 start-page: 774 year: 1996 end-page: 781 ident: BIB6 article-title: Effects of motion on performance of a pulse oximeter in volunteers publication-title: Anesthesiology contributor: fullname: Shah – volume: 11 start-page: 392 year: 1995 end-page: 393 ident: BIB3 article-title: Interference in a pulse oximeter from a nerve stimulator publication-title: J Clin Monit contributor: fullname: Stahl – volume: 4 start-page: 290 year: 1988 end-page: 301 ident: BIB4 article-title: Pulse oximetry: an analysis of theory, technology, and practice publication-title: J Clin Monit contributor: fullname: Peterson – volume: 83 start-page: 269 year: 1996 end-page: 272 ident: BIB5 article-title: Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room publication-title: Anesth Analg contributor: fullname: Tremper – volume: vol 25 start-page: 137 year: 1987 end-page: 154 ident: BIB2 article-title: Pulse oximetry: technical aspects of machine design publication-title: International Anesthesiology Clinics of North America contributor: fullname: Pologe – volume: 70 start-page: 98 year: 1989 ident: 10.1016/S0952-8180(97)00152-9_BIB1 article-title: Pulse oximetry publication-title: Anesthesiology doi: 10.1097/00000542-198901000-00019 contributor: fullname: Tremper – volume: vol 25 start-page: 137 year: 1987 ident: 10.1016/S0952-8180(97)00152-9_BIB2 article-title: Pulse oximetry: technical aspects of machine design contributor: fullname: Pologe – volume: 85 start-page: 774 year: 1996 ident: 10.1016/S0952-8180(97)00152-9_BIB6 article-title: Effects of motion on performance of a pulse oximeter in volunteers publication-title: Anesthesiology doi: 10.1097/00000542-199610000-00012 contributor: fullname: Barker – volume: 4 start-page: 290 year: 1988 ident: 10.1016/S0952-8180(97)00152-9_BIB4 article-title: Pulse oximetry: an analysis of theory, technology, and practice publication-title: J Clin Monit doi: 10.1007/BF01617328 contributor: fullname: Wukitsch – volume: 11 start-page: 392 year: 1995 ident: 10.1016/S0952-8180(97)00152-9_BIB3 article-title: Interference in a pulse oximeter from a nerve stimulator publication-title: J Clin Monit doi: 10.1007/BF01616745 contributor: fullname: Block – volume: 83 start-page: 269 year: 1996 ident: 10.1016/S0952-8180(97)00152-9_BIB5 article-title: Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room publication-title: Anesth Analg doi: 10.1213/00000539-199608000-00012 contributor: fullname: Dumas |
SSID | ssj0004521 |
Score | 1.5218155 |
Snippet | One of the sources of error in pulse oximetry readings is associated with an abnormal signal-to-noise ratio. The pulse oximeter distinguishes the light... |
SourceID | proquest crossref pubmed elsevier |
SourceType | Aggregation Database Index Database Publisher |
StartPage | 591 |
SubjectTerms | Electric Stimulation - adverse effects Electric Stimulation - instrumentation evoked potential stimulators Evoked Potentials, Somatosensory - physiology False Positive Reactions Humans Intraoperative Period Male Middle Aged nerve stimulators Oximetry - adverse effects Oximetry - instrumentation Oxygen - blood Peripheral Nerves - physiology pulse monitors |
Title | False low pulse oximetry reading associated with the concomitant use of a peripheral nerve stimulator and an evoked-potential stimulator |
URI | https://dx.doi.org/10.1016/S0952-8180(97)00152-9 https://www.ncbi.nlm.nih.gov/pubmed/9347439 https://search.proquest.com/docview/79366638 |
Volume | 9 |
hasFullText | 1 |
inHoldings | 1 |
isFullTextHit | |
isPrint | |
link | http://sdu.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV1Nb5wwELU2yaWXqlUbdfvpQw-tEOkCu2AfV82mSQ-9bCrlhoxtJNQNrHaXNPkH_dmdsQ2mrVb9kHoAIWQw4j3Gz8OMh5DXCU4SIqbDeCoVTFCUCLnQSZhOS65ggOXK-HTPl9mnK3a6mC5Go64SqD_3X5GGc4A1Zs7-Bdr9TeEEHAPmsAfUYf9HuJ9BDzpYNV-DdYtHzW11rXebu2Bjw-UD4QDp4s5RecKkGJ6lworCQWu9-wJXNK7MogOroMa4yACswTVW-3Jxl2AZ9E3zRatw3eww6MiknnRN9qjePhNTgI2FrreV8Da_UtYde7HynF1WygQczG9qH0j8AYsmgbB1lTpPK6wL5nxyyuX0ZS6Xr_erdbk1P4R-gvYDWx3ZOk8n2ppnliUh_ioe2m8-oGk2sMUzWwbMDeszWzj3lxHDOi-WfW-g63FtYo5iMg65Hyb74EXTFpvyzDTiB-QoBjMHVvZofrG4-jhcrT6yaz3ae_sMsne-wzc8e-s626eN9s19jAa6fEDuOxjp3LLuIRnp-hH5ZhhHgXHUMI52jKOOcdQzjiLjKMBOB4yjLV5UUkE946hhHPV0osA42OjPjBs0eUw-ny0u35-Hrr5HKJN0sgtjLUUB-jdKuYBZq4oLVfCJ4oXmMO8A6c_KOJUgoBKVlgodA5MyFjMpM55JkbDkmBzWTa2fECqimSynkyLSDCRpoYtSMsUy3MUyEmJMTro3m6_tMi65j28EKHKEIjfBGABFzseEde8_d1rUaswciPO7S191eOVgq_EHHHxSTbvNgSQpKHw2JscWxv5ZeDJFz8DTf-_1GbnnP63n5HC3afULcrBV7UtHy-_2QMK7 |
link.rule.ids | 315,782,786,27933,27934 |
linkProvider | Elsevier |
openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=False+low+pulse+oximetry+reading+associated+with+the+concomitant+use+of+a+peripheral+nerve+stimulator+and+an+evoked-potential+stimulator&rft.jtitle=Journal+of+clinical+anesthesia&rft.au=Keidan%2C+Ilan&rft.au=Sidi%2C+Avner&rft.au=Gravenstein%2C+Dietrich&rft.date=1997-11-01&rft.pub=Elsevier+Inc&rft.issn=0952-8180&rft.eissn=1873-4529&rft.volume=9&rft.issue=7&rft.spage=591&rft.epage=596&rft_id=info:doi/10.1016%2FS0952-8180%2897%2900152-9&rft.externalDocID=S0952818097001529 |
thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0952-8180&client=summon |
thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0952-8180&client=summon |
thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0952-8180&client=summon |